Owner's manual

56
Technical specifications
Eppendorf SOP
English (EN)
9.4.3 Multipette stream / Multipette Xstream
Combitip
advanced
Volume range Testing
volume
Error limits
Error
Systematic
error
Random
error
± % ± µL ± % ± µL
0.1 mL
(white piston)
Increment: 0.1 µL
1 µL - 100 µL 10 µL ±1.6 ±0.16 ±2.5 ±0.25
50 µL ±1.0 ±0.5 ±1.5 ±0.75
100 µL ±1.0 ±1.0 ±0.5 ±0.5
0.2 mL
(blue piston)
Increment: 0.2 µL
2 µL - 200 µL 20 µL ±1.3 ±0.26 ±1.5 ±0.3
100 µL ±1.0 ±1.0 ±1.0 ±1.0
200 µL ±1.0 ±2.0 ±0.5 ±1.0
0.5 mL
Increment: 0.5 µL
5 µL - 500 µL 50 µL ±0.9 ±0.45 ±0.8 ±0.4
250 µL ±0.9 ±2.25 ±0.5 ±1.25
500 µL ±0.9 ±4.5 ±0.3 ±1.5
1 mL
Increment: 1 µL
10 µL – 1 000 µL 100 µL ±0.9 ±0.9 ±0.55 ±0.55
500 µL ±0.6 ±3.0 ±0.3 ±1.5
1,000 µL ±0.6 ±6.0 ±0.2 ±2.0
2.5 mL
Increment: 2.5 µ
25 µL – 2 500 µL 250 µL ±0.8 ±2.0 ±0.45 ±1.125
1,250 µL ±0.5 ±6.25 ±0.3 ±3.75
2,500 µL ±0.5 ±12.5 ±0.15 ±3.75
5 mL
Increment: 5 µL
50 µL - 5 000 µL 500 µL ±0.8 ±4.0 ±0.35 ±1.75
2,500 µL ±0.5 ±12.5 ±0.25 ±6.25
5,000 µL ±0.5 ±25 ±0.15 ±7.5
10 mL
Increment: 10 µL
0.1 mL – 10 mL 1 mL ±0.5 ±0.005 ±0.25 ±0.0025
5 mL ±0.4 ±0.02 ±0.25 ±0.0125
10 mL ±0.4 ±0.04 ±0.15 ±0.015
25 mL
(blue adapter)
Increment: 25 µL
0.25 mL - 25 mL 2.5 mL ±0.3 ±0.0075 ±0.35 ±0.0088
12.5 mL ±0.3 ±0.0375 ±0.25 ±0.0313
25 mL ±0.3 ±0.075 ±0.15 ±0.0375
50 mL
(dark gray adapter)
Increment: 50 µL
0.5 mL - 50 mL 5 mL ±0.3 ±0.015 ±0.50 ±0.025
25 mL ±0.3 ±0.075 ±0.20 ±0.05
50 mL ±0.3 ±0.15 ±0.15 ±0.075