Technical specifications
Eppendorf SOP
English (EN)
9.6.1 Air-cushion pipettes, fixed and adjustable volume
9.6.2 Positive displacement pipettes (Biomaster)
Nominal volume Error limits
Measurement errors
Systematic Random
± % ± µL ± % ± µL
1 µl ±5.0 % ±0.05 µl ±5.0 % ±0.05 µl
2 µl ±4.0 % ±0.08 µl ±2.0 % ±0.04 µl
5 µl ±2.5 % ±0,125 µL ±1.5 % ±0.075 µl
10 µl ±1.2 % ±0,12 µL ± 0.8 % ±0.08 µl
20 µl ±1.0 % ±0.2 µl ±0.5 % ±0.1 µl
50 µl ±1.0 % ±0.5 µl ±0.4 % ±0.2 µl
100 µl ± 0.8 % ±0.8 µl ±0.3 % ±0.3 µl
200 µL ± 0.8 % ±1,6 µL ±0.3 % ±0.6 µl
500 µl ± 0.8 % ±4,0 µL ±0.3 % ±1.5 µl
1 000 µL ± 0.8 % ±8,0 µL ±0.3 % ±3,0 µL
2 000 µL ± 0.8 % ±16,0 µL ±0.3 % ±6,0 µL
5 000 µL ± 0.8 % ±40,0 µL ±0.3 % ±15,0 µL
10 000 µL ±0.6 % ±60,0 µL ±0.3 % ±30,0 µL
For Multi-channel pipettes, the error values are two times the values
specified for single-channel pipettes.
Nominal volume Error limits
Measurement errors
Systematic Random
± % ± µL ± % ± µL
5 µl ±2.5 % ±0,13 µL ±1.5 % ±0.08 µl
10 µl ±2.0 % ±0.2 µl ±1.0 % ±0.1 µl
20 µl ±2.0 % ±0.4 µl ± 0.8 % ±0,16 µL
50 µl ±1.4 % ±0,7 µL ±0.6 % ±0.3 µl
100 µl ±1.5 % ±1.5 µl ±0.6 % ±0.6 µl
200 µL ±1.5 % ±3.0 µL ±0.4 % ±0.8 µl
500 µl ±1.2 % ±6.0 µL ±0.4 % ±2.0 µL
1 000 µL ±1.2 % ±12.0 µL ±0.4 % ±4.0 µL