Limitation
The test strips are used for fresh capillary whole blood samples.
1. DO NOT use neonate blood sample.
2. Extreme humidity may affect the results. A relative humidity greater than 90may cause incorrect results.
3. The system should be used at a temperatures between 50 and 104℉℉ (10 and 40). Outside this
range, the system may get incorrect results.
4. DO NOT reuse the test strips. The test strips are for single use only.
5. Hematocrit: The hematocrits between 20% and 60% will not significantly affect the results. Hematocrit
below 20% may cause higher results. Hematocrit above 60% may cause lower results. If you do not know
your hematocrit level, please consult with your healthcare professional.
6. Altitude up to 10000 feet above sea level has no effect on readings.
Healthcare Professionals – Please note these additional Limitations
7. Patients undergoing oxygen therapy may have inaccurate results.
8. If the patient has the following conditions, the result may fails:
Severe dehydration
Severe hypotension (low blood pressure)
Shock
A state of hypoglycemic-hyperosmolar state (with or without ketosis)
9. Lipemic samples: Cholesterol level up to 500 mg/dL and triglycerides up to 3,000 mg/dL do not affect the
results. Grossly lipemic patient samples have not been tested and are not recommended for testing with
EasyPlus mini R2N Glucose Meter.