5. Squeeze the bottle until a
tiny drop forms at the tip of
the control solution cap.
6. Touch the drop to the
Blood Collection Area at
the end of the test strip.
7. Do not put control solution
on top of the test strip.
8. When you hear the beeper, you have
enough control solution in the test strip.
The meter starts to count down from 5
seconds and will show the results.
9. A result appears on the display. Don’t
remove the test strip yet. Check if the
reading falls within the range printed on
the test strip bottle.
10. Remove the test strip and
throw it away after you have
compared the reading to the
range printed on the test
strip bottle.