System information

Blastware Compliance Module
___________________________________________________
4–118 Blastware Operator Manual
(13) Exit
Click on Exit to leave the Auto Call Home program.
There are three communication options within Auto Call Home to remotely access data from the
monitor, 1) Serial (Communications), and 2) UDP/IP, and 3) TCP/IP.
i Once all setups for Auto Call Home are completed, it is strongly recommended that
communications be checked before the unit is placed in the field. Some modem/phone
combinations may require parameter changes to function properly. Refer to Chapter 6 for
more information on setting up modems and communications settings.
ii Auto Call Home is supported by Blastmate III or Minimate Plus units with a serial
number of 6000 or greater.
c. Status
Select the Status to access the Event Copy status function.
(1) Event Copy
Click on Event Copy to access the Event Copy Status dialog box. This window displays the
status of each communications Port, the Serial Number of the unit associated with that Port,
and the time and date of each connection.
d. Test
Use the Test feature of Auto Call Home to determine if any of the Events that have been copied
were created due to a false trigger. You may also email the files, using the parameters set in
Blastware Mail.
Sample of a Valid Event –the waveform signature is almost equivalent in the positive and
negative baseline.