System information
Blastware Compliance Module
___________________________________________________
4–46 Blastware Operator Manual
To choose a recording rate, click on the Sample Rate drop–down menu. Click the mouse pointer
on the required Sample Rate.
(6) Auxiliary Trigger
The Auxiliary Trigger is an external triggering feature used to start monitor recording.
An optional Wire Break/Make Trigger Cable is available (Part No. 720A3501) used to connect
several monitors together for multiple point blast monitoring.
To use the Auxiliary Trigger, set the monitor in Waveform Manual, Waveform, Histogram, or
Histogram Combo Record Modes. Set a fixed Record Time. The Auxiliary Trigger combined
with the Auto Record Time setting is only available on Minimate Pro units. Click on the
Auxiliary Trigger checkbox to display a check mark.
(7) Edit Compliance Setup More
i Geophone Alarm Level
The alarm level controls the triggering of an optional Remote Alarm (Part No. 720A3901).
The remote alarm can sound two alarms: warning and alarm. The warning alarm sounds when
Event activity exceeds the Trigger Level for the Geophone. The alarm level can be set to a
different level that is higher than the trigger level of the Geophone. This allows for the triggering
of a second alarm at the higher level.
For example, in pile driving operations the remote alarm triggers a warning when activity exceeds
the trigger level for the Geophone, notifying the operator that the pile driving activity has
exceeded that set level. The second alarm triggers when activity exceeds the alarm level, set to a
higher level, notifying the operator to re-evaluate the piling operation.
To set an alarm level, double click on the Alarm Level box. Enter the new level that is within the
stated range (Range: 0.508 – 254 mm/s).
ii Microphone Alarm Level
For a description of the optional remote alarm, see the Geophone channel in the previous section.
To set an alarm level, click on the Microphone Alarm Level box. Enter the new level that is
within the stated range (Range: 100.0 – 148 dB(L)).
iii Auxiliary I/O
The Auxiliary I/O section of the dialog controls the option for triggering of a Remote Alarm
(Part No. 711A1401) based on levels set for Vibration or Pressure, or for an External Trigger,
usi
ng the optional Wire Break/Make Trigger Cable (Part No. 720A3501). Select Off if you
don’t want to use the Auxiliary I/O option.