DR. RAJENDRA PRASAD CENTRE FOR OPHTHALMIC SCIENCES ALL INDIA INSTITUTE OF MEDICAL SCIENCES Ansari Nagar, New Delhi – 110029 NOTICE INVITING TENDER Stores Officer, Dr. R.P. Centre on behalf of Director AIIMS invites sealed tenders in two bid systems from the reputed firms/manufacturers for the purchase of following items up to 4.30P.M on or before 25/10/2012 :File No.
36/U-3/RPC/SSK/12-13 37/U-3/RPC/SSK/12-13 38/U-3/RPC/SSK/12-13 39/U-3/RPC/SSK/12-13 40/U-3/RPC/SSK/12-13 41/U-4/RPC/SSK/12-13 42/U-4-OT/RPC/SSK/12-13 43/U-4-OT/RPC/SSK/12-13 44/U-4/RPC/SSK/12-13 1000/500/500/1000/500/1000/1000/100/- 40000/35000/18000/30000/20000/30000/30000/10000/- 1000/100/100/100/100/1000/100/100/100/100/1000/- 40000/10000/5000/5000/5000/40000/10000/6000/6000/6000/40000/- 55/U-6/RPC/SSK/12-13 Wave front diagnostics and topolyzer – 01 No Ophthalmic refraction Unit – 16 Nos.
84/EB/RPC/SSK/12-13 85/ Path/RPC/SSK/12-13 86/ Path/RPC/SSK/12-13 87/ Path/RPC/SSK/12-13 88/ Path/RPC/SSK/12-13 89/ Path/RPC/SSK/12-13 90/ Path/RPC/SSK/12-13 91/ Path/RPC/SSK/12-13 92/ Path/RPC/SSK/12-13 93/ Path/RPC/SSK/12-13 94/ Path/RPC/SSK/12-13 95/ Path/RPC/SSK/12-13 96/ Path/RPC/SSK/12-13 97/Bio/RPC/SSK/12-13 98/Bio/RPC/SSK/12-13 99/Bio/RPC/SSK/12-13 100/Bio/RPC/SSK/12-13 101/Bio/RPC/SSK/12-13 102/Bio/RPC/SSK/12-13 103/Bio/RPC/SSK/12-13 104/Bio/RPC/SSK/12-13 105/Bio/RPC/SSK/12-13 106/Bio/RPC/SSK/12-13
138/COD/RPC/SSK/12-13 139/COD/RPC/SSK/12-13 140/UPS/RPC/SSK/12-13 141/UPS/RPC/SSK/12-13 Camera SLR D90 – 01 Nos. Refraction set – 03 Nos. Stand-by Back-up Module compatible with existing H/W (For 2 year Rate Contract for Different Centre of AIIMS) – 100 Nos.
Rs. /- DR. RAJENDRA PRASAD CENTRE FOR OPHTHALMIC SCIENCES ALL INDIA INSTITUTE OF MEDICAL SCIENCES ANSARI NAGAR: NEW DELHI: 110 029 (SCHEDULE-‘A’) S.NO. OF TENDER : _______ FILE NO. : Tender No. Name of the party in whose Favour the Tender form has been issued : _____________________________________________________ _____________________________________________________ _____________________________________________________ _____________________________________________________ The Chief, Dr. R.P.
CHECK LIST FOR TERMS AND CONDITIONS Check list for Terms and Conditions (To be filled by the bidder and submitted alongwith the technical bid) 1. Certificate for being in business for more than 2 years________________________________ 2. Certificate for sole ownership/partnership _________________________________________ 3. Statement of financial standing from bankers _______________________________________ 4.
ANNEXURE – I DR. RAJENDRA PRASAD CENTRE FOR OPHTHALMIC SCIENCES ALL INDIA INSTITUTE OF MEDICAL SCIENCES ANSARI NAGAR: NEW DELHI : 110 029 (SCHEDULE-‘B’) Tender Ref. No. : Subject Date of Submission Date of Opening : Purchase of : On or before 29-10-2012 upto 4.30.P.M : F. No. 1/U-1&2 to 68/U-6:- 30/10/2012 at 11.00 AM : F. No. 69/Casualty to 120/Pharma: - 31/10/2012 at 11.00 AM : F. No. 121/Anaes to 141/UPS: - 01/11/2012 at 11.00 AM /RPC/SSK/12-13 1. Tender should be addressed to the Chief, Dr. R.P.
:: 2 :: 5. The tenderers should give rates, showing taxes, if any, and levies, packing forwarding and insurance charges separately giving full breakup details. THE INSTITUTE IS NOT AUTHORIZED TO ISSUE ‘C/D FORMS’. PLEASE EXCLUDE CUSTOM DUTY COMPONENT IN RUPEE QUOTE AS THE INSTITUE IS EXEMPTED FROM THE PAYMENT OF CUSTOMS DULY. However, exercise Duty, if any, should be shown separately. Tender not confirming to this requirement shall be rejected and no correspondence will be entertained whatsoever. 6.
:: 3 :: 18. EACH TENDER SHOULD BE ACCOMPANIED WITH AN EMD/BID SECURITY AMOUNTING TO Rs._________ (as indicated in the tender Notice)/- BY WAY OF DEMAND DRAFT/ BANK GUARANTEE DRAWN IN FAVOUR OF “AO, Dr. R.P. Centre, AIIMS, NEW DELHI”, (PREFERABLY BANK GUARANTEE) FAILING WHICH THE TENDER SHALL NOT BE CONSIDERED FOR ACCEPTANCE AND WILL BE OUTRIGHTLY REJECTED.
:: 4 :: 35. Genuine equipments and instruments etc., should be supplied. Tenderers should indicate the source of supply i.e. name & address of the manufacturers from whom the items are to be imported, country of origin, country of Shipment etc. 36. The quoted equipment should preferably be of Indian/International standards. 37. The tenderers are required to quote the mode of shipment by Air/Sea/ Airport Parcel and should give separate breakup of freight and Insurance Charges. 38.
:: 5:: 45 The tenderer shall furnish a non-blacklisting certificate that the firm has not been blacklisted in the past by any government/Private institution. The tenderer/supplier has to give an affidavit on non-judicial stamp paper of Rs.10/that there is no vigilance/CBI case pending against the firm/supplier and the firm has not been blacklisted in the past by any Govt. or Private Organization. 46 Payment of Agency Commission, if any, payable in rupees must be indicated.
SECTION – XV LIST OF SCHEDULED BANKS A. Nationalized Banks 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. Allahabad Bank Andhra Bank Bank of Baroda Bank of India Bank of Maharastra Canara Bank Central Bank of India Corporation Bank Dena Bank Indian Bank Indian Overseas Bank Newbank of India Oriental Bank of Commerce Punjab National Bank Punjab & Sindh Bank Syndicate Bank Union Bank UCOBank United Bank of India Vijaya Bank B. State Bank of India and its Associated Banks 21. 22. 23.
F.No.1/U-1&2/RPC/SSK/12-13 90D Lenses – 6 Nos.
F.No. 2/U-1&2//RPC/SSK/12-13 28D Lenses- 6 Nos.
F. No. 3/U-1&2/RPC/SSK/12-13 ECG Machine – 1 Nos.
F. No. 4/U-1&2/RPC/SSK/12-13 B-Scan Probe 10MHz – 1 Nos.
F. No.5/U-1&2/RPC/SSK/12-13 Ophthalmic Chair without panel – 2 Nos.
F. No. 6/U-1&2/RPC/SSK/12-13 Ophthalmic Chair panel – 1 Nos.
F. No. 7/U-1&2/RPC/SSK/12-13 Pulse Oxymeter – 1Nos.
F. No. 8/U-1&2/RPC/SSK/12-13 Adjustable table for USG - 1Nos.
F. No. 9/U-1&2/RPC/SSK/12-13 Desktop Video Magnifier - 1Nos.
F. No.
F. No.
F. No. 13/U-1&2/RPC/SSK/12-13 Cardiac Monitor – 1 Nos.
F.No. 14/U-1&2/RPC/SSK/12-13 Endo Laser 532 -3 Nos 1. Should be a diode pumped frequency doubled solid state laser 2. Should have a 532 mm operating wavelength 3. Should have a forced air cooling system 4. Should have a aiming laser wavelength of 635 nm 5. Should have a out put power ranging from 30 mW to 2000 mW (2W) 6. Should have a exposure time of 10ms – 2000 ms and continous wave 7. Should be compatible with Endo laser probes 8. Should have provision to connect two doctors filter 9.
F. No. 15/U-1&2/RPC/SSK/12-13 Indirect Ophthalmoscope – 13 Nos • Binocular Indirect6 Ophthalmoscope with precision viewing upto 1.0 mm pupil size. • Spot size: 3 integrated spot size small spot, medium spot and large spot. • Filters: 4 integrated filters to choose from red filter, cobalt blue filter, yellow filter and diffuser. • Vertical adjustment, +/- 4° • Integrated flip up adjustment optics which can be flipped and locked at 0°, 12.5°, 47.5°, 60°.
F. No. 16/U-1&2/RPC/SSK/12-13 Binocular Video Indirect Ophthalmoscope – 6 Nos • High resolution A-Cam camera, CCD 470,000 Pixel, 460 lines. • Image sensor. ½” CCD , color • Fully integrated camera system. No external brackets or prisms. • 100% dustproof system. Can be placed in disinfectant solution (but cannot be autoclaved) • Optics specially- developed. Maximum brightness. • Focus adjustable for any working distance. From 250 to 800 mm. • Automatic white balance.
F. No.
F. No. 18/U-1&2/RPC/SSK/12-13 Xenon with accessories/ Xenon Light Source- 4 Nos Equipment having facility of Xenon Vapour illumination and laser from laser console to provide Coaxial laser and illumination through single fibre optic cable. Increasing illumination level possible without increasing photo toxic risks. 1. Improved Xenon Vapour illumination with two illumination ports. 2. Suitable for 20ga, 23ga, 25ga, 27ga & 29ga products 3. Facility of coaxial laser and illumination through one fibre. 4.
F. No. 19/U-1&2/RPC/SSK/12-13 Perkins Hand held Tonometer – 6 Nos • Hand Held Applanation tonoter using Goldmann prism, with precisely counter balanced movements for use in supine and sitting patients.
F. No. 20/U-1&2/RPC/SSK/12-13 Transequator Lens – 8 Nos Primary Application-Mid Peripheral Retinal Diagnosis and Focal/Grid Laser Therapy • Widefield of view past the equator for pan retinal imaging and treatment • Excellent substitute for Rodenstock pan fundus Lens Field of View 110°/132° Image Mag. .7x Laser Spot Mag. 1.
F.No. 21/U-1&2/RPC/SSK/12-13 Iredotomy Lens - 4 Nos A 10mm diameter, A 66D magnifying button in the anterior surface of the lens is positioned over the peripheral iris to give a clear view of the iridectomy site. Laser efficiency is increased compared with using no lens. The lens also helps stabilize the patient’s eye and retains the eye lids. Laser Spot Mag. Contact Diam. Lens Height Product Code Image Mag. OAIA 1.5x .67x 15.5mm 16.
F. No. 22/U-1&2/RPC/SSK/12-13 Capsulotomy Lens – 4 Nos Stabilizes the patient’s eye and minimizes the possibility of pitting the IOL during Nd: YAG laser capsulotomy. A 10mm diameter, 66D magnifying button in the centre of the lens enhances visualization and allows precise laser focus on the posterior capsule. Product Code Image Mag. Laser Spot Mag. Contact Diam. Lens Height OAYA 1.8x .56x 15.5mm 16.
F. No. 23/U-1&2/RPC/SSK/12-13 Imaging System for Fundus Camera & Slit Lamps– 2 Nos Camera System: • E6 High-Resolution 6MP Imaging System: • 6MP Film Size • Color/Monochrome Digital Sensor • Imaging System • Angiography Software • PC computer system (as per Sonomet Escalon Specifications) • Fire Wire Interface and Cables PC System: • Pentium Core 2 QUAD 2.
F.No. 24/U-1&2/RPC/SSK/12-13 Surgeon Chairs (Imported)– 12 Nos Anatomically formed back rest to reduce fresher intervertibral discs and provide support for lumbar spine. Seat system to evenly distribute surgeon weight. Forward leaning seat to support proper blood circulation and reduced fatigue. Arm Rest : Sturdy, height adjustable and fully movable in all directions. Seat System : Flat Chasis : • Width 60cm • Depth 60cm • Height 95.
F. No. 25/U-1&2/RPC/SSK/12-13 OT Table– 12 Nos Chasis: • Frame profile 90 X 50 mm • Outside length 805 mm • Outside width 605 mm • Central/broke on both sides, working on four wheels. The direction of two wheels can be independently adjusted. • Double rimmed casters.
F. No.
F. No.
F. No. 28/U-3/RPC/SSK/12-13 Keratoprosthesis – 100 Nos. The Boston Keratoprosthesis type I, machined from medical grade polymethylmethacrylate (PMMA) which is a collar button design keratoprosthesis .It should be composed of a front plate with a stem, which houses the optical portion of the device, a back plate and a titanium locking c-ring. Both the single standard pseudophakic power and customized aphakic optic with an 8.5 mm diameter adult size or 7.0 mm diameter pediatric size back plate are required.
F. No. 29/U-3/RPC/SSK/12-13 Endothelial donor corneal punches Disposable , single ready to use donor corneal punches with the cutting block base having four holes in the base of the block and four steel guide posts at the edges for corneal transplantation with trephine diameters as follows : Numbers required 8mm diameter-100 7.5mm diameter-100 7.25mm diameter-100 7.
F. No. 30/U-3/RPC/SSK/12-13 Hessberg barron suction trephines – 100 Nos. Single use vacuum trephine for partial penetrating keratoplasty with the suction system with disposable syringe with plunger. The trephine blad should be superfine quality. The sample should be given by the firms and the selection will be based on technical performance. Trephine diameters required: 7.0mm-100 7.25 mm- 100 7.5mm – 100 8.
F.No. 31/U-3/RPC/SSK/12-13 Lamellar dissectors'- 12 Nos. 1. Short, triangled spatula used to dissect downward into the cornea 2. Half-length, curved spatula with a rounded tip used to make the dissection up to two thirds across the cornea. 3. Full-length, curved spatula used to complete the dissection across the cornea over 360 degrees.
F. No. 32/U-3/RPC/SSK/12-13 Tear Osmometre – 01 No. Tear osmometer to measure the tear osmolarity should be able to measure osmolarity between 280 to 350 mosm/kg H2O with a standard deviation of 5 mosm/kg H2O and resolution of 1 mosm/kg H2O .
F.No. 33/U-3/RPC/SSK/12-13 Corneal Asthesiometer – 01 No. The corneal Aeshesiometer for rapid corneal sensitivity examinations Should be based upon the principle of pressure transmitted axially by a nylon monofilament, of known diameter but of variable length, for a given bending stress. The length of the thread like mono-filament, controlled by the forefinger should range from 60mm to 5mm. A scale on the Aesthesiometer instrument should indicate the length of the monofilament.
F.No. 34/U-3/RPC/SSK/12-13 Auto Non Contact Tonometer and Non Contact Pachymeter – 16 Nos. Fully integrated tonometer + pachymeter, combining the measurement of IOP and CCT into one instrument. Specifications: Tonometer Range : 0-60 mmHg Pachymeter Range: 200-900 micrometers With color LCD screen and printer.
F.No. 35/U-3/RPC/SSK/12-13 Autoref with Keratometer- 16 Nos. Objective and subjective mode and measuring corneal astigmatism, low contrast glare acuity testing. Measureable range-sphere plus/minus 20D, Cyl 0 to 7D, Axis 0 to 180, min. pupil size 2mm, vertex dist. 10.5, 12.0, 13.5, preferably with IOL mode and print out facility. High accuracy measurements of corneal and contact lens radii and determination of corneal astigmatism. Distance independent co-independent measuring technique.
F.No. 36/U-3/RPC/SSK/12-13 Wave front diagnostics and topolyzer – 01 No Wave front diagnostic device works on Tshering principle. It should be compatible with the Wavelight excimer laser system. The resolution should be at least 13 by 13 dots. The measuring beam should be diode 660nm and should allow pupil size to measure 3 to 8 mm. It should be able to measure – 12 to +6 D sphere and -6 to + 6D astigmatism. The centration should be done by eye tracking function.
F.No. 37/U-3/RPC/SSK/12-13 Ophthalmic refraction Unit – 16 Nos. 1. One fully upholstered comfortable ophthalmic chair with facility of full motorized recline and up & down movements for 300 mm ± 15 mm 2. One stand and console with illuminating light for examination. 3. The strand should have adequate space for placing Keratomter or autorefractomter, NCT, Lensometer, Direct Ophthalmoscope and streak retinoscope, Chart projector & Trial lens set.
F.No. 38/U-3/RPC/SSK/12-13 Slit lamp – 03 Nos. 1. Should have illumination form top using tungsten filament bulb capable of giving illumination intensity up to 600,00Lux. 2. Magnification from 6.3Xto 40X having continous zoom 3. Slit length 0.2 to 8mm 4. Slit length 0.2 to 8mm 5. Eyepieces of 12.5X 6. Facility to tilt the slit image up to 20 deg. Should be available 7. Should have applanation tonometer 8.
F.No. 39/U-3/RPC/SSK/12-13 Autolens Analyser – 08 Nos. Auto Lens Analyzer for accurate determination of power of various lenses including prisms, multifocals, CL and high index lenses; marking centration facility; Abbe value; software to adjust for different lenses including integrated spectrometer with measurement of UVA UVB with visible light transmission qualities of the lens.
F.No. 40/U-3/RPC/SSK/12-13 Artificial anterior chamber – 100 Nos. Single use disposable artificial anterior chamber with three pieces: base with tissue pedestal, tissue retainer and locking ring. The base should have provision for silicone tubing, in-line pinch clamps and female Luer Lok connectors. The port should allow injection of viscoelastic , balanced salt solution or air beneath donor cornea. The firm should submit the sample and selection will be based on technical performance.
F.No. 41/U-4/RPC/SSK/12-13 Hand held kerotometer – 01 No. Radius Curvature: Range: 5.00 – 10.0 mm Step : 0.01 mm Refractive Power : Range : 33.75 – 67.50 D (n=1.3375) Step : 0.01 mm Refractive Power : Range : 33.75 – 67.50 D (n=1.3375) Step : 0.01/ 0.12 / 0.25 D Anstigmatism : Range : 0 - ± 10.00 D Steps : 0.01 / 0.12 / 0.25 D Axis: Range : 0 - 180° Step : 1° Measuring Area : 3.3 mm (R=7.7 mm) Eccentricity : Range : 4.10 - ± 2.
F. No. 42/U-4 OT/RPC/SSK/12-13 USG A&B Scan – 01 No.
F. No. 43/U-4 OT/RPC/SSK/12-13 Hand held slit lamp – 01 No. • • • • • Slit size adjustable from 0.2mm x 10mm to 4mm x 14mm 6x magnification Blue interference filter (FITC), can be selected for corneal examination. Multi-coated optics for maximum light transmission. 2.5V or 3.5V XHL Xenon Halogen Technology for bright, white light comparable with the brightness of a classic slit lamp • A&B Scan Network and Connectivity Five USB 20.
F. No. 44/U-4/RPC/SSK/12-13 Laser Machine for selective Laser Trabeculoplasty – 01 No. Laser System : Portable Q-Switched, frequency-doubled Nd:YAG;532 nm wavelength Pulse Duration : 3 Nanoseconds Repetition Rate : 3 Hz Energy Settings : 0.30 to 2.0 mJ Pulse Setting : Single pulse-on-demand Spot Size : 400 micron Aiming Beam : Diode laser; coaxial, 635 nm wavelength, variably intensity Laser Delivery : Clip On Type; compatible with slit lamps Display : Remote Control. Should show exact no.
F. No. 45/U-4/RPC/SSK/12-13 Fibre Optic probe for canaloplasty – 20 Nos.
F. No. 46/U-4/RPC/SSK/12-13 6 Mirror Gonio Lens – 04 Nos. Field of View – 6X63” Image Mag – 1.0x Laser Spot Mag – 1.0x Contact Diameter – 8.
F. No. 47/U-4/RPC/SSK/12-13 4 Mirror hand held Gonioscope Lens – 04 Nos. Four 64° mirrors for complete anterior chamber angle viewing with minimal lens rotation. Directly hand held for easy handling and stability . Choice of large or small holding ring. Small diameter contact surface allows static or dynamic goioscopy without methylcellulose. Advanced technology, multi layer polymer coating protects mirrors and is compatible with most disinfecting methods.
F. No. 48/U-4/RPC/SSK/12-13 Hoskins Barkan Goniotomy Lenses – 02 Sets Designed for transverse goniotomy surgery with the operating microscope, but can also be used as a diagnostic lens. The infant lens is oval and conical in shape, with a 10mm diameter magnified view of the anterior chamber and anterior chamber angle. The premature infant lens is the same in shape and design except the dimension are 1 mm smaller for premature infant surgery. An adult size of 11.5 mm diameter is also available.
F. No. 49/U-4/RPC/SSK/12-13 UBM with B-Scan – 01 Nos. A. Transducer Frequencies: B Scan Probe : 10 MHz 20 MHz UBM Probe 35 MHz 50 MHz B. Probe characteristics (10MHz): Lateral Resolution : 0.0963 cm +/ 10% Axial Resolution :+ 0.190 mm+/-10% Focal Length : 13 – 35 mm (Focal Point 20 mm ± 2m) Maximum Frame Rate : 10 FPS Maximum Scan Angle : 60° Depth of Field : 80 mm @ 0.5 magnification Probe characteristics (20 MHz): Lateral Resulution : 0.0784 cm+/-10% Axial Resolution : 0.
B Scan : Scan Rate: 10FPS Scan Angle : 60° +/-0.5% Scan Depth:60 mm +/- 10% E. Display Monitor / Resolution: Display Monitor Maximum Resolution : 1280 X 1024 UBM Image Depth / Width : 512 x 1024 B Scan Image Depth/Width: 512 x 1024 UBM Display in grayscale and color palette. 3 F. Display Functions: Multiple Calipers, including Arbitrary A-Scan. This unique function provides a very precise tool to measure distances on microns scale.
K. Human Interface Devices: Keyboard Mouse/Trackball Combo Three-button Footswitch, which permits complete control of all the functions meanwhile the user keeps both the hands occupied on the exam L. AC Input Voltage: 90-132VAC or 180-264VAC Scan Conversion Maximum Frames per second: 22 FPS Vector per Frame: 256 Rays Samples per Vector: 1024 O.
F. No. 50/U-4/RPC/SSK/12-13 Non-Contact Tonometer – 01 No. Height : 1975 in (50.2 cm) Width : 10.5 in (26.7 cm) Depth : 14 in (35.6 cm) Weight : 23 1bs (10.
F. No. 51/U-4/RPC/SSK/12-13 Collagen Implant for Trabeculectomy – 50 Nos. A biodegradable and implantable scafforld matrix, inducing a regenerative non-scarring would healing process for implantation under the conjunctiva during trabeculectomy.
F. No. 52/U-4/RPC/SSK/12-13 Drainage device for glaucoma surgery – 20 Nos. A stainless steel drainage device for use in glaucoma surgery To shunt aqueous from the anterior chamber to the subconjunctival space External diameter 30-500 um Internal diameter 30 – 100 um Lenth 1.
F. No. 53/U-4/RPC/SSK/12-13 Glaucoma Drainage Implant with valve – 20 Nos. A silicone material implant with plate and tube for glaucoma surgery Plate/Valve Plate Specifications: Thickness: 0.9-1.2 mm Width: 13.00mm-15 mm Length: 16.00mm-18 mm Surface Area: 184.00-200 mm 2 Tube Specifications: Length: 25.00-30.00mm Inner Diameter: 0.305-0.40 mm Other Diameter: 0.635-.
F. No. 54/U-6/RPC/SSK/12-13 Anterior segment oncular photography system with Video capture card • • • • • • Video & still camera facilities for high quality digital images Video capture card.
F. No. 55/U-6/RPC/SSK/12-13 Perkins Hand held applanation tonometer – 06 Nos. Hand held applanation tonometer using Goldman prism, with precisely counter balanced movements for use in supine and sitting patients. Battery based handle. Additional rechargeable handle with recharging unit.
F. No. 56/U-6/RPC/SSK/12-13 Tonometer- Tonographer – 01 No. Instrument to provide ocular pulse amplitude measurements, in addition to tonometry and a tonography mode to measure the facility of aqueous outflow.
F. No.
F. No. 58/U-6/RPC/SSK/12-13 MRI film scanner- 3 Nos. + 01 illumination Box Scanning Area: 2.5" x 2.5" (Min.) - 14" x 35" (Max.
F. No. 59/U-6/RPC/SSK/12-13 Distance Rondot Steroacuity test – 03 Nos Test for distance stereopsis for patients of ages 4 years and above.
F. No. 60/U-6/RPC/SSK/12-13 Portable Autorefractometer – 01 No. Light Weight, hand held Measurement Range S -18D to +23D C 0 to - 12D, 0 to +12D Axis: 1° to 180° Increments Auto/0.25D selectable Axis: 1° Measurement Area 2.5mm -Consistency Value 10 points -Reaching Distance 10.9 inch. (278mm) -Working Distance 2 inch.
F. No. 61/U-6/RPC/SSK/12-13 Focimeter – 01 No. For accurate analysis of the power of any lens including multifocal, prisms, contact lens, centration with software to adjust for all lens materials including integrated spectrometer for quick, presice measurement for UVA and UVB and invisible light quality of lens. Specifications: Sphere Measuring range-25+25D unit 0.01/0.12/0.25D Cylinder Measuring range-9.99+9.99D unit0.01/0.12/0.25D Axis Measuring range 0180 unit1 Add Measuring range0+9.99D unit0.01/0.12/0.
F. No. 62/U-6/RPC/SSK/12-13 Non invansive continouse IOP monitoring device – 02 Nos. Proprietary items of the Sensimed triggerfish Consists of soft hydrophilic contact lens with a guage embedded in the silicone and an antenna. The antenna is attached to a cord to which is connected to a portable recorder. The patient wears the contact lens 24 hours including during sleep.
F. No. 64/U-6/RPC/SSK/12-13 Hand Held digital fundus camera with anterior segment attachment and Laptop – 01 No Non mydriatic fundus camera light weight portable with anterior segment photography attachment. The smartscope by Optomed is one such device.
F. No. 65/U-6/RPC/SSK/12-13 Perimetric progression software – 01 No PROGRESSOR software It is a software that analyses visual field progression in glaucoma patients. Provides a bar chart for each test location.
F. No. 67/U-6/RPC/SSK/12-13 Goldmann Projection type Kinetic primeter – 01 No. Stimulus size I.V, Goldmann intensities 1a to 4e, with manual stimulus movement.
F. No. 68/U-6/RPC/SSK/12-13 Confocal Scanning Laser opthalmascopy (HRT III) – 01 No. The HRT3 is the only imaging device with distinct databases for individual ethnic groups, Glaucoma Probability Score, or GPS, is a new approach to analyzing HRT data using the latest artificial intelligence methods, producing an easy way to identify the probability of disease. Topographic Change Analysis (TCA) is a statistically based progression algorithm that accurately detects structural change over time.
F. No. 69/Casualty/RPC/SSK/12-13 Electric Hydraulic Ophthalmology Operation Table (Imported) 5 Nos. ¾ ¾ ¾ ¾ ¾ Multi purpose function for Ophthalmology operation and examination use. Disinfectant Resistant stainless steel finished. Special design hand holder, more convenience for doctors in operation. All stainless steel accessories including clamps. Safety Backup all functions or movement of the unit will stop automatically by relief valve. While the maximum ranges are reached.
F. No. 70/Casualty/RPC/SSK/12-13 Nd. Yag Laser 1 Nos. • • • • • Laser wavelength 1064nm, Structure Mode: super-Gaussian for highly precise beam profile. Optical breakdown 2.5 Mj or less in air Pulse duration < 4ns Max. Laser energy 10mj ( Single Pulse), 23mj(Double pulse) And 37mj (Triple pulse) • • • • • • Minimum Energy 0.3Mj – 10mj( Single Pulse) Energy levels:22 steps Pulse repetition frequency Max.2 Hz. Focus diameter 10 micron in air Cone angle/Angle of exit aperture 16 Deg.
F. No. 71/Casualty/RPC/SSK/12-13 A & B Scan (Ultra Sound) – 01 No. Fully automated A / B Scan with Biometry facility, should have video printer. Specification for the UltraScan A / B : B–Scan:.Specification The unit should have a Sealed Probe , 10Mhz, for taking the B-Scan Image. It should have facility for documenting the image on a video printer. Should provide standard video output in RS-170,S-video,RGB. This unit should have a minimum of 256 shades of gray for enhanced Resolution.
Slit Image radial range: +/- 90 degrees Radial movement of slit light illumination relative to microscopic axis: horizontal +/- 90 degrees, vertical 0-20 degrees Filters: blue, red free, grey 10%, heat-absorbing filter Light source: Light-emitting diode Working Temperature: 0-50 degrees Objective lens focal length: 105 mm Magnification changer: 6.3 x, 10x, 16x, 25x, 40 x Ocular magnification: 12.
F. No. 73/OPD/RPC/SSK/12-13 Hand-held pediatric autoref – 02 Nos. Objective should measure corneal astigmatism. Measurable rang sphere ≥ plus/ minus 20D, Cyl ≥ 10D, Axis 0 to 180, min. pupil size 2.0, 2.3mm, vertex dist. 10.5, 12.0, 13.5, preferably with IOL mode and print out facility. Light in weight, operable with one hand.
F. No. 74/OPD/RPC/SSK/12-13 Slit Lamp (with provision of digital camera attached) – 01 No. 1. Should have illumination from top using tungsten bulb capable of giving illumination intensity up to 600,000 Lux. LED powered. Lifetime usage, no replacement required with tilt facility. 2. Magnification from 6.2X2.40Xhaving continous zoom having steps at 10.16, 25,40X 3. Diameter of field should be from 32mm to 5.1mm 4. Slit Length 0.2-8mm 5. Eyepiece of 12.5x 6.
F. No. 75/OPD/RPC/SSK/12-13 Non Contact tonometer with pachymeter – 08 Nos.
F. No. 76/EB/RPC/SSK/12-13 Class 2 Biosafety Cabinet – 01 No. The equipment should have following features: • Cabinet should be of Class II type A2, provided with stand. • Normal Size of Biosafety Cabinet: 1.2 meters (4 feet).
F. No. 77/EB/RPC/SSK/12-13 Slit lamp with camera – 01 No. The equipment should have following features: • Background Illumination adapter • Beam Splitter • Digital Adapter • Camera 12.2 mpx • Software for Photo Management Canon • Power adapter • Memory Card 8 GB • Electric Table • Camera parameters are setup, so you don't have to figure out the correct settings.
Chin-Rest: Vertical movement : 80mm Fixation : Green LED Power: Input voltage : 220V/110V~±10% Input frequency : 50Hz/60Hz Power consumption : 30VA(max) Output voltage : Lamp 6V (continuously adjustable) Fixation 3V Dimension & Weight: Dimension : 740mm x 450mm x 500mm Gross weight : 25Kg Net weight : 24kg Working environment: Temperature : +5~+40 Relative humidity : ≤80% Air pressure : 800hpa~1060hpa Storing environment: Temperature : -40~+55 Relative humidity : ≤93% Air pressur
F. No. 78/EB/RPC/SSK/12-13 Digital camera with a special lenses – 01 No. The equipment should have following features: Type Digital, single-lens reflex, AF/AE camera Recording Media SD memory card, SDHC memory card, SDXC Memory Card Image Format 22.3mm x 14.9mm (APS-C size) Compatible Lenses EF lenses including EF-S lenses (35mm-equivalent focal length is approx. 1.6x the lens focal length) Pixels Effective pixels: Approx. 18.
1/4000 sec. to 30 sec., bulb (Total shutter speed range. Available range varies by shooting mode) Soft-touch electromagnetic release 10-sec. or 2-sec. delay or 10-sec. delay plus continuous shooting Remote Switch RS-60E3 Shutter Speeds 1/4000 sec. to 1/60 sec., X-sync at 1/200 sec. 1/4000 sec. to 30 sec., bulb (Total shutter speed range. Available range varies by shooting mode) Self Timer 10-sec. or 2-sec. delay or 10-sec.
F. No. 79/EB/RPC/SSK/12-13 Hand held slit lamp – 01 No. The equipment should have following features: • • • • • • • Rechargeable, cordless power supply for slit-lamp examination anywhere. Compact and lightweight, so that the examiner can hand hold the unit over a long period with minimal fatigue. Extra-bright halogen lamp light source. Easy, simple selection of one of three slit widths or spot illumination. Quick, one-touch selection of 6x or 10x magnification. Built-in blue filter.
F. No. 80/EB/RPC/SSK/12-13 Eye Bank specular Microscope – 01 No. The equipment should have following features: 0.48X0.60mm Viewing Field 0.20 X 0.
F. No. 81/EB/RPC/SSK/12-13 Eye Retrieval kit – 20 Nos.
F. No. 82/EB/RPC/SSK/12-13 Air purifier – 01 No. The equipment should have following features: • Electronic controls and 4 fan speeds • Turbo setting • Should have Sleep mode facility • True HEPA filter to capture 99.97% of particles as small as 0.
F. No. 83/EB/RPC/SSK/12-13 UV lighter - 04 Nos.
F. No. 84/EB/RPC/SSK/12-13 Binocular self illuminated loupe – 02 Nos. The equipment should have following features: • It should have illumination from a 6v 20w Halogen Lamp. • Built in blue and red free filters & three step field of illumination. • Compact and light weight for longer periods without fatigue. • Precision optics and elegant mechanism, synchronizes the convergence and parallel adjustments. • Condensing lens & carrying case, Teaching Mirror, S. Depressor, Funds chart, spare Bulb.
F. No. 85/Path/RPC/SSK/12-13 Cryostat – 01 No. - The Instrument should be free-standing motorized open top with independent specimen temperature control via Two separate compressor system. - Convenient section thickness selection from outside the cryo-chamber. - Section thickness range : 0.5 to 300 um. - Temperature setting range –10 deg. To –50 deg. C. - Cryo-chamber temperature setting down upto –40 deg. C. - Specimen retraction : 50 um.
F. No. 86/Path/RPC/SSK/12-13 Nano-Drop spectrophotometer – 01 No. - Micro-volume sample size (0.5-2.
F. No. 87/Path/RPC/SSK/12-13 Tissue processing machine – 01 No. - Carousel type with 12 stations of 1.8 litre each;10 reagent stations, 2 wax baths. - Metal tissue basket having less Base diameter compare to upper diameter 200, cassettes capacity. - Delayed start function up to 9 days. - 9 freely selectable programme.
F. No. 88/Path/RPC/SSK/12-13 Fluorescent microscope – 01 No. - Microscope stand with harmonic component system and variable Koehler illumination. - Transmitted light with built-in stabilized power supply ,Halogen lamp12V-100W Three Step focusing. - 7 fold objective reversed nosepiece. - Trinocular tube with 25/30 degree viewing angle and three way beam splitting with 100/50/0. - Rotatable mechanical stage with ultra hard ceramic stage plate and coaxial drive. - Eyepiece 10x/22,focusable.
F. No.
F. No. 90/Path/RPC/SSK/12-13 Laminar air flow – 01 No.
F. No. 91/Path/RPC/SSK/12-13 Real Time PCR machine – 01 No. - Thermal Cycling System- Pelter based System - Block format- 48 well - Supported volumes- 10-30 µl - Supported consumables -48 well (0.1ml) plates with optical adhesive covers - 48 well (0.1ml) plates with optical flat caps - 8 tubes (0.1ml) strips with optical flat caps - Individual (0.1 ml) tubes with optical flat caps - Sample ramp rate – Fast mode: +/- 2.2 ºC/sec - Standard mode: +/- 1.6ºC/sec - Peak block ramp rate – 4.
F. No. 92/Path/RPC/SSK/12-13 Electrophorisis Power Supply – 01 No - Out put type : constant voltage or constant current with automatic cross over - 0utput voltage: 10-300VDC - Output current : 4 to 400mA - Increments : 1mA - Timer: 0 to 999 minutes - Display : 3 digit LED - Output terminals : Four - Dimensions : 13x8.7x2.9 in ./33x22x7.4cm - Weight:1.9Kg/4.
F. No. 93/Path/RPC/SSK/12-13 Bench TOP pH Meter – 01 Nos. - Bench top PH meter UB-5:Ph/temperature, - Simple one button standerization with ph range 0.00-14.00. - Auto buffer recoganization from 3 sets of 16 buffers, resolution 0.001 ph, accuracy 0.005 ph, - Automatic temperature compensation and temperature range 0.0-100 C , - Resolution 0.1 C, Accuracy 0.2 C, - Unit Dimensions (LXWXH); 9.0X4.8X3.
F. No. 94/Path/RPC/SSK/12-13 Ultra Sensitive Weighing Balance – 01 Nos. - Readable to 0.0001gram - Auto-stability indicator - Real time temperature compensation - optional,motor driven, built-in calibration weight - a windscreen that measures 7.5Wx8.25Dx9.5H (inches) - a circular, 3.5inch weighing pan and standard RS232 interface - Capacity- 210g - Readability- 0.
F. No. 95/Path/RPC/SSK/12-13 Vacuum Drier (Centrifuge Evaporator) 01 Nos. - Centrifugal speed :1800 rpm(400xg) - Temperature range: 30 to 60oC - Lowest vacuum: 5mbar. - Dimensions:12.8x23x14H in - Integrated centrifugal concentrator with Teflon - Membrane pump and rotor for 24x1.5ml/48x.
F. No. 96/Path/RPC/SSK/12-13 Flow Cytometer – 01 No. - Air Cooled Argon ion Laser (488nm) and Red Diode Laser (635nm) - The working principle should be single platform, fluorescence based flow cytometry. - Capable of minimum 5 colour analysis. - Have user – interchangeable optical filters - Should have high performance forward scatter, side scatter and fluorescent detectors. - Should be compatible to all routinely used flurochromes and upgradable in future.
F. No. 97/Bio/RPC/SSK/12-13 Multimode Microplate Reader – 01 No 1. System should have a Dual Monochromator based Multi detection microplate reader with five following detection modes: (a). Fluorescence Intensity (Top & Bottom) (b). TRF (c). HTRF (d). UV-Vis Absorbance (e). Fluorescence’s Polarization (f). Glow Luminescence detection 2. General Photometric Performance (a). Plate formats : 6, 12, 24, 48, 96, 384 wells and Cuvette Ports Luminescence Mode (b).
(e). Sensitivity : ≤ 1 pM fluorescein in 96 wells < 1.5 pM in 384 wells. 5. Time Resolved Fluorescence (a). (b). (c). (d). (e). Reading capabilities : Top or Bottom of a Microplate Wavelength range : 250–850nm Wavelength selection : Monochromators, tunable in 1nm increments Bandwidth (EX,EM) : 9nm, 15nm Precision Data Collection : 1–100 flashes, delay of 0-600 µ.sec before integration time selectable between 50 -1500 µ.sec. (f). Sensitivity : 10 fM europium in 96 or 384 wells with top-read. read, 5.
¾ ¾ ¾ ¾ ¾ ¾ both distributor and manufacturer will be black listed for forever for supplying any kind of instruments to the Government’s Research Organization, Hospitals, Laboratories etc. in any where in India. Relevant literature and publications that support the quoted model's ability to perform all of the above capabilities must be attached.
F. No. 98/Bio/RPC/SSK/12-13 Refrigerated Ultracentrifuge - 01 No. Technical Specifications 1. Maximum Speed : 1,40,000 rpm or better 2. Maximum RCF : 10,00,000 g or better 3. Noise Level : 50 db or better 4. Speed Control Accuracy : ±50 rpm or better 5. Display : Touch sensitive LCD or better technology 6. Rotor Temperature Control : 0 – 40 0C or better 7. Timer: 1min to 80 hrs or better & 50 min with hold function with Real Time Control function 8. Heat radiation rate : 1 kw or better 9.
regulations of AIIMS. And both distributor and manufacturer will be black listed for forever for supplying any kind of instruments to the Government’s Research Organization, Hospitals, Laboratories etc. in any where in India. ¾ The entire tender specifications claim about the instrument quoted need to be supported with technical specifications published in manufacturer’s technical bulletin/technical notes/technical brochure etc.
F. No. 99/Bio/RPC/SSK/12-13 Ultra Micro Weighing Balance – 01 No. 1. Capacity x Readability : 2.1 g x 0.1 µg 2. Repeatability : ± 0.25 µg 3. Linearity : ± 0.4 µg 4. Response Time : 10 seconds 5. Tare Range : 2.1 grams 6. Sensitivity Drift : ± 1 ppm / °C 7.
F. No. 100/Bio/RPC/SSK/12-13 CO2 Incubator – 01 No. 1. Capacity: 184L 2. CO2 Concentration Range: 0 to 20% 3. O2 Range: 1 to 20% 4. Relative Humidity : Ambient to 95% at 37°C 5. Humidity System: 3L stainless-steel pan 6. Temperature Range : Ambient +5°C to 55°C 7. Temperature Uniformity: ±0.2°C at 37°C 8. Chamber : Polished stainless-steel interior with coved corners 9. Interior D x W x H: 20.0 x 21.3 x 26.8 in. 10. Shelving : Stainless steel 11. Exterior L x W x H: 25 x 26.0 x 39.5 in. 12.
F. No. 101/Bio/RPC/SSK/12-13 Deep Freezer -20°C – 01 No. 1. Temp. Range : -20°C 2. Capacity : 820 L 3. Door type: Double Door 4. System should have Insulated inner doors 5. System should have easy to use door lock, should have heated vacuum valve to prevent a vacuum when opening the door at short intervals 6. System should have Castors and door key lock 7. System should have Heated door frame for easy opening of door 8. System should have Micro processor controller with digital display 9.
F. No. 102/Bio/RPC/SSK/12-13 Deep Freezer -90°C – 01 No. 1. System should have Dual Refrigeration System and two independent compressors both units are controlled independently and the two separate cooling system 2. Temp. Range : -90°C 3. Capacity : 820 L 4. Door type: Double Door 5. System should have Insulated inner doors 6. System should have easy to use door lock, should have heated vacuum valve to prevent a vacuum when opening the door at short intervals 7.
F. No. 103/Bio/RPC/SSK/12-13 Digital Dry Bath Incubator – 01 No. 1. Temp. Control Range : 5 to 100 0C 2. Timing Range : 1min to 99hr 3. Temp. Control Accuracy: ≤±0.3 4. Display Accuracy : 0.1 5. Heating Time: ≤12min 6. Cooling Time : ≤30min 7. Heating & Cooling : Peltier type 8. Should have anodized aluminum block 9. Should have places for 24 tubes x 1.5ml 10. Should have precise microprocessor control 11. Should have programmable temperature and timer 12. Should have wide temperature setting range 13.
F. No. 104/Bio/RPC/SSK/12-13 Ice Flaking Machine – 01 No. 1. Ice flake production capacity : 25 to 300 Kg Per Day 2. Refrigeration : CFC / HCFC Free 3. Collecting Bin Capacity : 25 to 45 Kg 4. Interior construction : 304 Stainless Steel 5. Outer construction : 20 SWG CRCA Sheet 6. Ice production Start Time : 10-15 Minutes 7. Should have micro processor based temperature control system 8. Should have in-built Auto Cut off 9. Should have castor wheels for easy mobility 10.
F. No. 105/Bio/RPC/SSK/12-13 Multi Channel Electronic Pipette – 01 No. 1. Volume range: 0.5-10µl, 5-100µl, 15-300µl, 50-1000µl 2. Systematic error : ±2.0 to ± 5.0 % or better 3. Random error: ±0.2 to ± 0.8 % or better 4.
F. No. 106/Bio/RPC/SSK/12-13 Sonicator – 01 No. 1. Power Rating: 150W 2. Variable power output: 0-150 watts 3. Output Frequency: 20-25 KHz 4. Sample volume: 10-100 mL 5. Standard Titanium Horn: Ø3, Ø2, Ø6 6. Should have Digital display, Fully microprocessor controlled and completely programmable 7. Should have Auto-tuning for convenience of use and optimal processing efficiency 8.
F. No. 107/Bio/RPC/SSK/12-13 Analytical Wieghing Balance – 01 No. 1. Readability : 0.01mg 2. Weighing capacity : 220 g 3. Average stabilization time : 2 sec 4. Average response time : 6 sec 5. Repeatability [mg] : 0-20 g: 0.015 6. Linearity : 0.1 7. Minimum sample weight : 0.
F. No. 108/Bio/RPC/SSK/12-13 Automatic Safety Burner – 01 No. Safety Bunsen burner with following details : 1. Display movement sensor 2. Flame monitor 3. Overheating protection 4. 3 operation modes : Manual, Infrared sensor, Foot switch 5. 2 nozzles for propane/ butane and natural gas 6. Adapter for standard gas hose with inner diameter 10 mm 7. Main Adapter 8.
F. No. 109/Bio/RPC/SSK/12-13 Atuomatic Digital Desiccators – 01 Nos. 1. Stainless steel and glass 2. Superior seal integrity 3. Rigid door full length door hinges 4. One piece neoprene door gasket 5. Shelves adjustable in half-inch increments 6. Able to Automatically regenerate the silica gel desiccant at every 5 hours 7. Easy to read door mounted hygrometer 8. Overall W x L x H: 20 1/2" x 23" x 22 1/4") 9.
F. No. 110/Micro/RPC/SSK/12-13 Gel Documentation and Image Analysis System (Imported) – 01 No. 1. GLP compliant Fully Automated Microprocessor based computer controlled Gel Documentation & Image analysis system 2. CCD Camera with 16-bit performance (not through file generation) and motorized zoom lens (10 X) with close-up facility. 3. Fully enclosed camera/lens assembly for minimizing the risk of cross contamination of hazardous chemicals between users 4.
F. No. 111/Micro/RPC/SSK/12-13 70°C Deep freezer (Imported) – 01 No - Vertical -86oC freezer to operate on 220 V/50Hz and ambient temperature of 20oC - Capacity 400 liters or above. - Stainless steel non corrosive chamber with atleast 4 racks which could be closed separately independent of the main door.
F. No. 112/Micro/RPC/SSK/12-13 Tissue Homogenizer (Imported) – 01 No. 1. Homogenizer for homogenizing tissue/ cells. 2. Should be ideal for homogenizing, dispersing and emulsifying samples in aqueous media, and tissues. 3. Should be a light weight, hand held and also stand mountable, portable unit. 4. Should be able to work with low as well as high volumes (0.25 to 30 ml) and (50 to 500ml). 5. Should have adjustable speed from 3,000 rpm to 30,000 rpm. 6.
F. No. 113/Pharma/RPC/SSK/12-13 Co2 Incubator – 01 No. • • Interiors must be made up of stainless steel or copper The internal volume of the incubator should range between 200 – 250 L. • It must have minimum 6 door gas tight segmented access. • It should have dry heat sterilization option for decontamination.
F. No. 114/Pharma/RPC/SSK/12-13 Cold Centrifuge • Maximum speed in RCF 20000 g • The centrifuge must be able to accommodate tubes from 0.
F. No. 115/Pharma/RPC/SSK/12-13 Rheometer • Speed range 0.01-1200 RPM. • Torque range 0.25-27 mNm. • Shear Stress – 0.5 to 3 x 10 4 Pa. • Shear Rate range 10-2 to 4000. • Viscosity measuring range from 1 to 109 mPas. • Automatic temperature measurement 20-180oC.
F. No. 116/Pharma/RPC/SSK/12-13 Benchtop NMR spectrometer – 01 No. • The instrument should have liquid-phase proton NMR system. • It should have permanent magnet, transmitter, receiver, data acquisition and programmable pulse sequencer. • It should have Ethernet interface and intuitive web-based control software. • The larmor frequency should range between 40 to 50 MHz. • Resolution should be better than 100 ppb. • SNR should be 300 for water, single shot.
F. No. 117/Pharma/RPC/SSK/12-13 Multi Mode Microplate Reader • The Instrument should be a spectral scanning multimode reader including absorbance, fluorescence intensity (top/bottom reading), photometric and ultraluminometric detection technologies and supports endpoint, kinetic and spectral scanning measurements. • Advanced Monochromator for example dual excitation monochromators and dual emission monochromator for all possible applications.
F. No.
F. No. 119/Pharma/RPC/SSK/12-13 Horizontal Gel electrophoresis – 01 No. 1. A horizontal electrophoresis system should be able to run the gel size of 7x10cm & 7 x 7cm in the same tank with safety lid. 2. A system should include UV transparent gel trays of size 7x10 & 7x7cm with integrated fluorescent ruler. 3. A system should include tape free gel casting module for leak free operations. 4.
F. No. 120/Pharma/RPC/SSK/12-13 2D Gel Electrophoresis with imaging System – 01 No. Specifications Complete system to separate, quantitate, characterize, visualize and analyze proteins in biological samples. System should have following features: A) Isoelectric focusing unit (1st Dimension): • • • • • • • • • • • System should include 12 different Individual Lane Control for running different samples, pH Gradients and focusing protocols in a single run with different protocols.
• Graphic LED display. Gel imaging system • System with true 16 bit CCD with more than 4 Megapixel image resolution. • Cooled CCD with cooling range of -30°C absolute temperature using peltier cooling system. (if any other technology used, please quote its superiority) • System with dynamic flat fielding technology for getting accurate Quantitation with CV < 5 %. • Multi image color target with multi-channel image display for study of more than 2 dyes at a time for differential expression.
F. No. 121/Anaes/RPC/SSK/12-13 Videolaryngoscope - 01 No. 1. One complete set of 6 laryngoscope blade including Macintosh blades with metal finish size 2, 3 and 4. 2. Laryngoscope blade size 0 & 1 can be macintosh or miller blade. 3. One special blade for difficult intubation with the facility for introduction of suction catheter for size 1618 Fr. 4. All blades should be with integrated camera chip and LED light illumination for obtaining more than 50000 Lux of brightness. 5.
F. No. 122/Anaes/RPC/SSK/12-13 Non – Invasive Pulse co-oximeter – 04 Nos. 1. Equipment should have Display type: High resolution TFTdisplay 2. Display of Pulse signal strength, plethysmograph and SpO2 values, Pulse rate, Perfusion index & trend must be quoted with additional parameters Oxygen Content (SpOC), Carboxyhemoglobin (SpCO), Methemoglobin(SpMet) 3. System should be upgradable to acoustic respiration rate 4. Display of short trend of SPO2 or PR (selectable) along with plethysmograph 5.
F. No. 123/Anaes/RPC/SSK/12-13 C trash Laryngeal Mask Airway – 01 No. 1. Should have viewer comprising of high resolution color monitor, built in light source, USB port for transfer of images, charging stand. 2. Should be supplied with reusable stabiliser rod for ET Tube. 3. Should have the provision to use the Flexometallic ET Tube made up of 100% silicon of different sizes.
F. No. 124/Anaes/RPC/SSK/12-13 Flash Autoclave – 02 Nos. 1. It should be a fully automatic table top high pressure steam sterilizer with chamber volume of 20 L for sterilization of wrapped, unwrapped, hollow and porous materials and should have cycles for prions also. 2. It should have Integrated steam generator and vacuum pump. 3. It should be a B class & N class sterilizer with cycle time for wrapped instrument should be less than 25 minutes. 4.
F. No. 125/Anaes/RPC/SSK/12-13 View a vein – 01 No. 1. It should be infrared light to highlight hemoglobin 2. Portable and lightweight 3. It is as simple to use as pointing the device at an area of the skin and clicking to display the peripheral veins beneath. 4. It should illuminate a square shaped area. 5. The device should be non- contact. 6. No sterilization should be needed between uses. 7. It should run on a rechargeable battery, so it doesn’t need to be plugged in to an electrical outlet while use.
F. No. 126/Anaes/RPC/SSK/12-13 CPR manikin adult with feedback – 01 No. 1. Full Body adult male manikin.The anatomical features desirable for instruction should be as identified by the American Heart Association 2. It should have patented hygienic system facilities provides comfort and safety for the trainee and makes internal cleaning entirely unnecessary. 3.
21. It should have ECG control mode, which allows the instructor to control essential parameters like heart rate, heart rhythm (ECG), pulse rate, defibrillation chance, respiration rate, saturation, consciousness, blood pressure etc. 22. Should give detailed performance report in which all BLS & ACLS performances on manikin should be registered second by second. 23. It should have facility to print the report of the CPR performance. 24.
F. No. 127/Anaes/RPC/SSK/12-13 CPR Manikin pediatric with feedback – 01 No. 1. Full Body Child manikin.The anatomical features desirable for instruction should be as identified by the American Heart Association 2. It should have patented hygienic system facilities provides comfort and safety for the trainee and makes internal cleaning entirely unnecessary. 3. It has a foam filled body with no internal parts to break.
21. It should have ECG control mode, which allows the instructor to control essential parameters like heart rate, heart rhythm (ECG), pulse rate, defibrillation chance, respiration rate, saturation, consciousness, blood pressure etc. 22. Should give detailed performance report in which all BLS & ACLS performances on manikin should be registered second by second. 23. It should have facility to print the report of the CPR performance. 24.
F. No. 128Anaes/RPC/SSK/12-13 Pediatric intravenous arm Trainer – 02 Nos. • The IV Trainer for teaching the insertion of cannulas and catheters, the infusion of fluids and the injection of medication, blood sampling and pulse measurement should have following features: - hole closure after penetration, extending the lifetime of the arm. • It should have movable wrist and rotating arm ensure an absolutely realistic training product, especially when a catheter is placed on the back of the hand.
F. No. 129/Anaes/RPC/SSK/12-13 Adult intravenous arm Trainer – 02 Nos. • The IV Trainer for teaching the insertion of cannulas and catheters, the infusion of fluids and the injection of medication, blood sampling and pulse measurement should have following features: - hole closure after penetration, extending the lifetime of the arm. • It should have movable wrist and rotating arm ensure an absolutely realistic training product, especially when a catheter is placed on the back of the hand.
F. No. 130/Anaes/RPC/SSK/12-13 Drug storage cart – 04 Nos. 1. Flexible storage system for the many anaesthesia supplies, medication & instruments required in the operating room 2. Should be made of ABS, scratch and corrosive resistant cart, powder coat paint inside and out 3. 4. 5. 6. Should have 5 year warranty Stabilize system to prevent cart from tipping over Castor wheels should be not less than 5 inches.
F. No. 131/Anaes/RPC/SSK/12-13 Radiant warmer – 02 Nos. 1. Floor Model Radiant Warmer with Ceramic heating element(s) with high-grade steel safety guard. 2. Non-visible infra-red radiation, therefore no impairment of visual examination. 3. Functions: it should have self test. 4. It should have intensity selection from 2 to 30 mW/cm2 (depending on distance to the patient pad) in increments of 2 mW/cm2. 5. Intensity selection should display in mW/cm2. 6.
F. No. 132/Anaes/RPC/SSK/12-13 Defibrilator with CPR Feedback – 01 No. 1. The defibrillator should be latest, lightweight, small size with bright colored display 2. The defibrillator should be with Rectilinear Biphasic technology, with 3 wave- form display on screen with minimum 6.5 inches screen size. 3. Weight of defibrillator should not be more than 6.5 Kgs 4. It should display of both selected and delivered energy 5. It should have ability to energy selection from Paddles as well as unit. 6.
F. No. 133/Anaes/RPC/SSK/12-13 Rigid fiberoscope, pediatric – 01 No. 1. Rigid fiberoptic scope with malleable tip for superior access to navigate in small spaces 2. Should be able to accommodate any ETT size 2.5 to 5.5 3. Should have light source, handle & oxygen port. 4. Should have adjustable tube stop for sliding up and down the stylet which should be able to fix and hold ET tube in place 5.
F. No. 134/Radio/RPC/SSK/12-13 Portable ultrasound color Doppler machine – 01 No. A state of art fully digital, compact portable Colour Doppler Ultrasound Machine is required with following technical features for use in Dr. R.P. Centre. 1. It should be suitable for complete orbital & eye applications, vascular access and examination of other small parts. It should allow ophthalmic and general applications both in adults and children & also suitable for interventions. 2.
F. No. 135/COD/RPC/SSK/12-13 Portable Slit lamp – 02 Nos. • • • • • • • This is a slit lamp microscope, with adjustable slit sizes and spot illumination. It is totally portable, compact, and lightweight with a lithium battery as power supply, It permits long-term continuous lighting for approximately 2 hours with no external power required; large capacity battery. It has extra bright halogen lamp light source with quick, one-touch selection of 10X or 16X magnification. It has a built-in blue filter.
F. No. 136/COD/RPC/SSK/12-13 Fundus Camera – 01 Nos. 1. 12.3 megapixel camera 2. (Mydriatric / Non-Mydriatric) Retinal Imaging System with IMAGE net Software 3.
F. No. 137/COD/RPC/SSK/12-13 Digital Foci meter – 01 Nos. Reading system: Illuminated display Type: digital projection lensmeter Sphere: 0-±25D, 0.01/0.06/0.12/0.25D steps Cylinder: 0-±10D, 0.01/0.06/0.12/0.25D steps Cylinder axis: 0-180°, 1° steps Prism power Range: 0-5∆,0.5∆ steps; 0-20∆(with prism unit), 0.
F. No. 138/COD/RPC/SSK/12-13 Camera SLR D90 – 01 Nos. 1. SLR D90 with VR kit 2. 105mm F2.8 Lense Fitted with Infrared UV Barrierfilters 3. SB 700 Flashgun (UV Photography) 4.
F. No. 139/COD/RPC/SSK/12-13 Refraction set – 03 Nos. Spheres From ± 0.12 to ± 20.0 Cylinders From ± 0.12 to ± 6.0 Trial frame with occlude, pin hole etc.
F. No. 140/UPS/RPC/SSK/12-13 Stand-by Back-up Module compatible with existing H/W & S/W (For 2 year Rate Contract for Different Centre of AIIMS). The back-up module should be capable of working with the existing H/W & S/W. The responsibility of linking the existing H/W – S/W and consumables with the battery standby module shall rest on vendor.
F. No. 141/UPS/RPC/SSK/12-13 LIPS system for Microprocessor based system (Wire Interactive power supply system) (For 2 year Rate Contract for Different Centre of AIIMS) 1. UPS should have wide input voltage range to run on the mains mode. Batteries are used only when absolutely necessary thus maximizing available back-up time and extending battery life. Rating: 600VA and 1000 VA 2. Automatic voltage regulation (AVR) to delivers stable voltage and provide higher availability. 3.