Specifications
15. Operating the On-board Data Logger
Page 100 USN52 L Operating Manual
15.2 Setting Number of Files
Readings can not be stored in the Data Logger until the desired
number of files has been set and one file has been selected to
receive readings. Number of files can only be set when Data Log-
ger memory is clear.
Use the following procedure to set number of files:
1. Switch to the lower level menus:
or
2. Frame the column containing the Data Logger functions
shown at right:
or ,
If the FILE NUMBER function displays ALL CLEAR memory
is clear and the number of data files can be set. If memory is
not clear, see Section 15.8 for instructions on clearing memory.
3. Press
(descending) or (ascending) aligned with FILE
NUMBER to display the desired number of files. Set #01 = 1
file, Set #02 = 2 files, etc.
4. When the desired option is displayed press
simulta-
neously. Configuring is displayed briefly while memory is
being set up.
The number of files determines the number of available locations
in each file. Table 15-1 on the next page lists the number of loca-
tions per file for each option.
NOTE: Configuring memory automatically enables the fol-
lowing menu function settings:
• FILE MODE is set to THICKNESS
• MEASURE is set to 0 TO 1st
• GATE LOGIC is set to POSITIVE
• COPY MODE is set to DATA LOGGER
These default settings may be changed individually by the
normal procedure.
When configured, the menu will display:
FILE
NUMBER
LOCATION
1
VALUE
EMPTY
FILE MODE
THICKNESS
Recording starts in File 1, Location 1 automatically, unless another
location or file is selected.