Specifications

USN52 L Operating Manual Page 131
Data Set Storage: Up to 140 data sets including A-scan image and all instrument parameters; RECALL resets
instrument to stored parameters and displays stored A-scan (frozen) for review and print;
optional naming of data sets
Data Logger: Storage capacity of 2500 digital measurement values; sequential storage with clear, re-
view, and direct report features; configurable up to 99 files; readings per file determined by
number of files configured
Interface: RS-232C bi-directional interface for data transfer and remote control; printouts formatted
for Epson FX-850 dot matrix printer with Intelligent Serial Interface number 8148
Outputs: Gate threshold violation; digital thickness value via RS-232 interface, TOF analog, ampli-
tude analog, sync, TTL Go/No Go
Probe Connectors: BNC or Lemo, specified at time of order
Power Supply: AC or 6 D-size NiCad, lead acid, or alkaline batteries
Battery Life: 10 hours on a single charge with NiCad batteries
Dimensions: 9.8" W x 5.25" D x 5.75" H (248.9 mm x 133.4 mm x 146.1 mm)
Weight: 5.98 lbs (2.718 kg) total weight with NiCad batteries; 3.10 lbs (1.41 kg) without batteries
and detachable battery pack
Environmental Temperature: Operating temperature: 0° C to 55° C (-20° C to +70° C survivable)
Storage temperature: -40° C to +75° C
17. Technical Specifications