Specifications

USN52 L Operating Manual Page 23
S= Measurement between first two echoes in gate GATE LOGIC OFF and
#= “Held” measurement value (no echo in gate) MEASURE = MULTECHO and
H= Amplitude of highest echo in gate ANGLE = OFF
Measurement Line content will be: When these functions are set to:
Current Data Logger Information as follows:
S= Angle beam soundpath distance to discontinuity GATE LOGIC OFF and
P= Projection (surface) distance to discontinuity MEASURE OFF and
D= Depth of discontinuity relative to sound entry surface ANGLE OFF
#= All three values held when no signal is in the gate
LOC: Current file location where reading will be stored MEASURE OFF and
VAL: Contents of current file location (LOC:) GATE LOGIC OFF and
FILE MODE = THICKNESS and
COPY MODE = DATA LOGGER and
ANGLE = OFF
Enlarged Value Display: A value highlighted in the measurement line using
is displayed in this area. Scrolling out of the
measurement line disables the enlarged value display. When the angle beam (trigonometric) flaw location
function, ANGLE, is enabled, scrolling out of the measurement line automatically displays the highest
amplitude echo signal in the gate.
When the data logger is enabled the measurement value is displayed in this area.
Status Line: A variety of symbols may appear in this line to show the status of certain functions. Status line
indicators are illustrated and explained on the next page.
4. Front Panel of the USN52 L