Specifications

9. Calibrating the A-scan Display
Page 66 USN52 L Operating Manual
9.5 Calibration - Angle Beam Probe
General procedure:
NOTE: This procedure requires a test block, suitable for
use with angle beam probes, which will produce reference
echo signals at two different, known, soundpath distances
from the probe. Commonly used standard blocks include
IIW, DSC, DC, and Angle Beam Calibration Block.
1. Turn on the USN52 L and set initial instrument settings per
Section 9.2.
2. Set PULSER (PULS menu) to SINGLE:
3. Set RANGE (BASIC menu) to at least the maximum thick-
ness or distance to be displayed in the A-scan:
,
4. Couple the probe to the test block.
5. Adjust GAIN until reference echo signals are at their desired
amplitude:
,
6. Adjust ZERO (S-CAL menu) to place the leading edge of the
first (from left) reference echo signal at its correct horizontal
position:
7. Adjust MTL VEL (S-CAL menu) to place the leading edge of
the second reference echo signal at its correct horizontal
position:
,
NOTE: Factory setting of coarse velocity #2 is close to the
shear wave velocity in steel, as measured on typical steel
test blocks.
8. Repeat steps 6 and 7 until both reference echoes are cor-
rectly positioned.
Sample specific procedure:
NOTE: The following procedure is an example only. It re-
quires a steel, IIW Type 1 test block, couplant, and a single
element angle beam probe. The A-scan will be calibrated
to a depth range of 10 inches of steel.
1. Turn on the USN52 L and set initial instrument settings per
Section 9.2.
2. Select the PULS menu:
or
3. Set PULSER to SINGLE:
4. Select the BASIC menu: