37
5.7 Settings
From the Home Screen, choose
“Settings – Insulin Delivery Settings”
as shown in Figure 40 to change delivery
settings.
Figure 40
1
Maximum Basal Rate: Used to limit the
maximum basal rate that can be set in the
Basal Menu
2
Base Basal Rate: Used to enter the
Base Basal Rate (Chapter 5.2)
3
Temporary Basal Rate: Choose to
adjust settings for temporary basal rates.
a) Off: Turn off the Temporary Basal Rate
function. The corresponding Temporary
Basal Rate button will be greyed out.
b) U/hr: Temporary Basal Rate will be
displayed in U/hr.
c) %: Temporary Basal Rate will be
displayed as a % of the current basal rate.
Setting this value to the most used
basal rate will simplify creating Basal
Programs.
Note
This feature is used to limit the
maximum amount of bolus delivery
that can be administered by accident or
misoperation.
Note
By default, the Temporary Basal Rate
function is off. If the button is greyed
out, activate the function using the
settings menu (Chapter 5.7)
Note