System Instruction

Table Of Contents
46
Equil
®
Patch Insulin Pump System
in the settings menu (Section 6.7). Also,
every time you press the Enter (Select)
button, you will hear a beep from the
pump to confirm that you have pressed the
button. Each time you press the button,
the tone will ascend in pitch (up to 5, and
then repeating) to help you remember
the number of times you have pressed the
button.
3
After you have set the correct bolus
amount, wait for 3 seconds and the pump
will repeat the audible tones back for you
to confirm the correct number of bolus
increments. If the number of tones is
correct, press the pump button one last
time, and the bolus will be administered.
6.6 Canceling a Bolus
A bolus can be stopped during delivery.
This can be helpful, for example, if you
notice that the bolus size is incorrect, or if
your meal plans change.
When the bolus is being administered,
the Home Screen is displayed and bolus
information is shown in the Information
Display Area.
Press the Cancel button and a
confirmation window will appear. Press
OK to confirm. You can also cancel a bolus
by pressing and holding the Quick Bolus
button on the pump.
Cancelling a bolus delivery will
cancel the total bolus (Amount Now and
Extended amount).
Note
The maximum number of increments
for a Quick Bolus is 20.
Note
If there is no activity for 10 seconds,
the pump exits the Quick Bolus mode.
Note
If you would like to cancel, simply
wait 10 seconds for the system to exit
the Quick Bolus mode.
Note