Users Manual

80
Equil
®
Patch Insulin Pump System
Figure 100
3
After the test has been completed, slide
the test strip ejector to pop out the test
strip.
If the control solution results are outside
of the reference range:
Confirm you are matching the correct
range. Control Solution 1 results should be
matched to the CTRL1 range printed on the
test strip vial (or foil pouch).
Check the expiration date of the test
strip and control solution. Make sure that
the packages have not been opened for
more than 6 months. Discard any expired
test strips and control solution.
Confirm that you are testing within the
correct temperature range (15-30 °C).
Make sure that the test strip vial and
control solution bottle have been tightly
closed.
Make sure that you are using the correct
brand of control solution.
Make sure you are following the user
guide instructions properly.
After checking all of the conditions
above, repeat the quality control test with
a new test strip.
If the quality control test results are still
The control solution test results will
not be stored in your history and the
test result tags will be greyed out.
Note