User guide
24/25
888-24-304-G-00 rev. B • 10/14
ENGLISH
Maintenance & Safety
Recommended Periodic Inspection and Maintenance
Component Action How often By whom
UI, Ethernet, USB 
cables
Inspect for wear, pinching, bad 
connectors
Monthly Any user
Casters Inspect for wear and debris Monthly Any user
Maximum Load
Inspect to ensure that maximum 
recommended loads are not 
exceeded
Daily Any user
Power System SLA 
batteries
Replace lead acid batteries
1,2
It is recommended that a discharge test be conducted when cart is  rst 
purchased to determine runtime with computer equipment con gured on 
the cart. Change batteries when instructed by StyleLink battery management 
software. If StyleLink is not used, it is strongly recommended that a monthly 
discharge test be conducted to determine runtime loss. If runtime is less than 
50% of original, batteries must be changed.
IT Personnel
Power Module
Inspect for dust at intake, vacuum 
as required using a vacuum cleaner 
that DOES NOT generate ESD 
(Electrostatic Discharge)
Monthly IT Personnel
Battery harness
Inspect for wear, cracking, pinching, 
or other damage
Monthly IT Personnel
Coiled Cord
Inspect for wear, damage, or 
stretching. 
Weekly Any user
1. Please dispose of all batteries in accordance with local law
2. Always replace with similar size battery (33 A-h to replace 33 A-h) and always replace in pairs
Earth Bond Test: To ensure safety grounding between the power cord ground connection and any 
accessible metal parts on the cart, although not mandatory, the following test is recommended to be 
performed every two years using a calibrated medical device safety analyzer. The procedure is as follows:
1. Disconnect the cart from mains power (unplug cart from the wall).
2. Remove power from internal power system outlets by pressing the power button on the Power System 
Interface.
3. Set up Earth Bond Test per the medical device safety analyzer instructions.
4. Connect the Cart power cord to the medical device safety analyzer.
5. Remove cover to access Lift Tension Adjustment point.
6. Attach second medical safety device analyzer probe to Lift Tension Adjustment point on Cart.
7. Perform Test (25 amps).
8. Ensure Earth Bond is less than or equal to 0.2-ohms.
9. Remove medical device safety analyzer connections to Cart, replace cover over Lift Tension 
Adjustment point, return the Cart to service. 
Insulation Resistance Test: To ensure mains power lines are adequately insulated from earth ground, the 
following test, although not mandatory, is recommended to be performed annually using a calibrated 
medical device safety analyzer. The procedure is as follows:
1. Disconnect the cart from mains power (unplug cart from the wall).
2. Remove power from internal power system outlets by pressing the power button on the Power System 
Interface.
3. Set up the Insulation Resistance Test per the medical device safety analyzer instructions.
4. Connect the Cart power cord to the medical device safety analyzer.
5. Perform test (500 V).
6. Ensure Insulation Resistance is greater than or equal to 1 Mohms.
7. Remove medical device safety analyzer connections to Cart and return Cart to service.
3OHDVH5HDGDQG)ROORZ3URGXFW0DQXDO,QVWUXFWLRQV
THIS IS A POLARIZED CORD - HOSPITAL GRADE ONLY
127(*URXQGLQJ5HOLDELOLW\&DQ2QO\%H$FKLHYHG:KHQ
WKH(TXLSPHQW,V&RQQHFWHG7R$Q(TXLYDOHQW5HFHSWDFOH0DUNHG
³+263,7$/21/<´2U³+263,7$/*5$'(´
820-672
,QVSHFW&RUG7KRURXJKO\%HIRUH(DFK8VH'212786(,)'$0$*('
'R1RW3OXJ0RUH7KDQ63(&,),('180%(52):$776,QWR3RZHU6\VWHP
'R1RW5XQ&RUG7KURXJK'RRUZD\V+ROHVLQ&HLOLQJV:DOOVRU)ORRUV
)8//<,16(573OXJ,QWR2XWOHW
'R1RW5HPRYH%HQGRU0RGLI\$Q\0HWDO3URQJVRU3LQVRI&RUG
'R1RW8VH([FHVVLYH)RUFHWR0DNH&RQQHFWLRQV
.HHS$ZD\)URP:DWHU'212786(:+(1:(7
.HHS&KLOGUHQ$ZD\)URP&RUG
'R1RW3OXJ&RUGLQWR$Q([WHQVLRQ&RUG
$92,'29(5+($7,1*8QFRLO&RUGDQG'R1RW&RYHU,W:LWK$Q\0DWHULDO
'R1RW'ULYH'UDJRU3ODFH2EMHFWV2YHU&RUG'R1RW:DON2Q&RUG
*5$633/8*WR5(029()5202XWOHW'R1RW8QSOXJ%\3XOOLQJ2Q&RUG
DANGER!
ELECTRICAL CORDS CAN BE HAZARDOUS
Misuse Can Result in FIRE or DEATH by ELECTRICAL SHOCK.
LES CORDONS ELECTRIQUES PEUVENT ETRE DANGEREUX
Une mauvaise utilisation peut provoquer un INCENDIE ou la MORT par CHOC 
ELECTRIQUE.
820-672