User guide
23 of 23
888-24-310-G-00 rev. C • 10/14
ENGLISH
Maintenance & Safety
Ergotron Electromagnetic Guidance and Manufacturer’s Declaration
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Powered Computer Cart is intended for use in the electromagnetic environment speci ed 
below. The customer or the user of the Powered Computer Cart should assure that it is used in 
such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1 The Powered Computer Cart uses RF energy only for its 
internal function. Therefore, its RF emissions are very 
low and unlikely to cause any interference in nearby 
electronic equipment.
RF Emissions
CISPR 11
Class A The Powered Computer Cart is suitable for use in all 
establishments other than domestic and those directly 
connected to the public low-voltage power supply 
network that supplies buildings used for domestic 
purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage  uctuations/
 icker emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Powered Computer Cart is intended for use in the electromagnetic environment speci ed 
below. The customer or the user of the Powered Computer Cart should assure that it is used in such 
an environment.
Immunity Test IEC 60601
Test Level
Compliance 
Level
Electromagnetic environment 
- guidance
Electrostatic 
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies Floors should be wood, 
concrete, or ceramic tile. If  oors 
are covered with synthetic 
material, the relative humidity 
should be at least 30%
Electrical Fast 
Transient/Burst
IEC 61000-4-3
±2 kV for power supply lines
±1 kV for input/output lines
Complies Mains power quality should be 
that of a typical commercial or 
hospital environment.
Surge
IEC 61000-4-5
±1 kV di erential mode
±2 kV common mode
Complies Mains power quality should be 
that of a typical commercial or 
hospital environment
Voltage 
Dips, short 
interruptions, 
and voltage 
variations on 
power supply 
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 
seconds
Complies Mains power quality should be 
that of a typical commercial or 
hospital environment. 
Power 
Frequency 
(50/60 Hz) 
Magnetic Field
IEC 61000-4-8
3 A/m Complies Power frequency magnetic 
 elds should be at levels 
characteristic of a typical 
location in a typical commercial 
or hospital environment.
Note: U
T
 is the AC mains voltage prior to application of the test level