User guide
23/29
1 a
a
bc
2
b
b
g
c
e
h
a
d
f
3
10mm (3/8")
ENGLISH
888-24-319-G-00 rev. C • 12/14
Turn o all mounted equipment.
How to Change Power System Batteries
Turn power system o by holding down
the AC Outlet Power button for 1 - 3
seconds. Power light will shut o .
Turn both circuit breakers to "OFF".
Red
Red
Black
BATTERY
LEAD
Recycle battery or contact
Ergotron for proper battery
disposal guidelines.
Black
BATTERY REPLACEMENT
The power module is congured for two, 33 -Ah batteries.
Only the following batteries are compatible with this
system:
• B&B EP33-12 • SBS S-12330
• Longway 6FM33G • Vision 6FM33D
• Vision 6FM33HD-X • PBQ PBQ33-12L
• Werker WKA12-33C • Werker WKA12-35C
• Duracell Dura 12-35C
Installing batteries other than the 33 -Ah listed above will
void the product warranty. Failure to heed this warning
may result in severe damage to batteries, power module
and possible re hazard.
REMPLACEMENT DE LA BATTERIE
Le module d’alimentation est conguré pour deux batteries 33 A-h.
Seules les batteries suivantes sont compatibles avec ce système :
• B&B EP33-12 • SBS S-12330
• Longway 6FM33G • Vision 6FM33D
• Vision 6FM33HD-X • PBQ PBQ33-12L
• Werker WKA12-33C • Werker WKA12-35C
• Duracell Dura 12-35C
L’installation de batteries autres que les batteries 33 A-h citées
ci-dessus annulera la garantie du produit. Si vous ne vous conformez
pas à cette précaution d’utilisation, les batteries et le module
d’alimentation risquent d’être sérieusement endommagés, ce qui
pourrait provoquer un incendie.
WARNING
AVERTISSEMENT
822-386
www.ergotron.com
BATTERY
LEAD
WARNING: RISK OF ELECTRICAL DISCHARGE
• Do NOT swap battery cables, doing so will cause arcing and trip
the circuit breaker.
• Do not remove or replace the batteries while cart is located in
an oxygen rich or hazardous environment, arcing may occur and
cause combustion.
• Replace rst battery before removing second battery to reduce
the risk of cables touching terminals and causing arcing.
• Always replace both batteries!
• Replace rst battery before removing second battery.
• Replace batteries with same Amp/Hour rating batteries only.
• Only Ergotron-speci ed batteries may be used in the StyleView
Power System. Please call customer care for more details.
• Recycle battery or contact Ergotron for proper battery disposal
guidelines.
Caution: Remove Black (-) before removing Red (+).
Caution: Connect Red (+) before connecting Black (-).
Caution: Do NOT swap battery cables, doing so will cause sparking and trip the circuit breaker.
Disconnect Power System and Multiple Socket Outlet from
power source.










