User guide

23/29
1 a
a
bc
2
b
b
g
c
e
h
a
d
f
3
10mm (3/8")
ENGLISH
888-24-319-G-00 rev. C • 12/14
Turn o all mounted equipment.
How to Change Power System Batteries
Turn power system o by holding down
the AC Outlet Power button for 1 - 3
seconds. Power light will shut o .
Turn both circuit breakers to "OFF".
Red
Red
Black
BATTERY
LEAD
Recycle battery or contact
Ergotron for proper battery
disposal guidelines.
Black
BATTERY REPLACEMENT
The power module is congured for two, 33 -Ah batteries.
Only the following batteries are compatible with this
system:
• B&B EP33-12 • SBS S-12330
• Longway 6FM33G • Vision 6FM33D
• Vision 6FM33HD-X • PBQ PBQ33-12L
• Werker WKA12-33C • Werker WKA12-35C
• Duracell Dura 12-35C
Installing batteries other than the 33 -Ah listed above will
void the product warranty. Failure to heed this warning
may result in severe damage to batteries, power module
and possible re hazard.
REMPLACEMENT DE LA BATTERIE
Le module d’alimentation est conguré pour deux batteries 33 A-h.
Seules les batteries suivantes sont compatibles avec ce système :
• B&B EP33-12 • SBS S-12330
• Longway 6FM33G • Vision 6FM33D
• Vision 6FM33HD-X • PBQ PBQ33-12L
• Werker WKA12-33C • Werker WKA12-35C
• Duracell Dura 12-35C
Linstallation de batteries autres que les batteries 33 A-h citées
ci-dessus annulera la garantie du produit. Si vous ne vous conformez
pas à cette précaution d’utilisation, les batteries et le module
d’alimentation risquent d’être sérieusement endommagés, ce qui
pourrait provoquer un incendie.
WARNING
AVERTISSEMENT
822-386
www.ergotron.com
BATTERY
LEAD
WARNING: RISK OF ELECTRICAL DISCHARGE
• Do NOT swap battery cables, doing so will cause arcing and trip
the circuit breaker.
• Do not remove or replace the batteries while cart is located in
an oxygen rich or hazardous environment, arcing may occur and
cause combustion.
• Replace  rst battery before removing second battery to reduce
the risk of cables touching terminals and causing arcing.
• Always replace both batteries!
• Replace  rst battery before removing second battery.
• Replace batteries with same Amp/Hour rating batteries only.
• Only Ergotron-speci ed batteries may be used in the StyleView
Power System. Please call customer care for more details.
• Recycle battery or contact Ergotron for proper battery disposal
guidelines.
Caution: Remove Black (-) before removing Red (+).
Caution: Connect Red (+) before connecting Black (-).
Caution: Do NOT swap battery cables, doing so will cause sparking and trip the circuit breaker.
Disconnect Power System and Multiple Socket Outlet from
power source.