Maintenance & Safety
Recommended Periodic Inspection and Maintenance
Component Action How often By whom
UI, Ethernet, USB
cables
Inspect for wear, pinching, bad
connectors
Monthly Any user
Casters Inspect for wear and debris Monthly Any user
Maximum Load
Inspect to ensure that maximum
recommended loads are not
exceeded
Daily Any user
Power System SLA
batteries
Replace lead acid batteries
1,2
It is recommended that a discharge test be conducted when cart is rst
purchased to determine runtime with computer equipment con gured on
the cart. Change batteries when instructed by StyleLink battery management
software. If StyleLink is not used, it is strongly recommended that a monthly
discharge test be conducted to determine runtime loss. If runtime is less than
50% of original, batteries must be changed.
IT Personnel
Power Module
Inspect for dust at intake, vacuum
as required using a vacuum cleaner
that DOES NOT generate ESD
(Electrostatic Discharge)
Monthly IT Personnel
Battery harness
Inspect for wear, cracking, pinching,
or other damage
Monthly IT Personnel
Coiled Cord
Inspect for wear, damage, or
stretching.
Weekly Any user
1. Please dispose of all batteries in accordance with local law
2. Always replace with similar size battery (33 A-h to replace 33 A-h) and always replace in pairs
Earth Bond Test: To ensure safety grounding between the power cord ground connection and any
accessible metal parts on the cart, although not mandatory, the following test is recommended to be
performed every two years using a calibrated medical device safety analyzer. The procedure is as follows:
1. Disconnect the cart from mains power (unplug cart from the wall).
2. Remove power from internal power system outlets by pressing the power button on the Power System
Interface.
3. Set up Earth Bond Test per the medical device safety analyzer instructions.
4. Connect the Cart power cord to the medical device safety analyzer.
5. Remove cover to access Lift Tension Adjustment point.
6. Attach second medical safety device analyzer probe to Lift Tension Adjustment point on Cart.
7. Perform Test (25 amps).
8. Ensure Earth Bond is less than or equal to 0.2-ohms.
9. Remove medical device safety analyzer connections to Cart, replace cover over Lift Tension
Adjustment point, return the Cart to service.
Insulation Resistance Test: To ensure mains power lines are adequately insulated from earth ground, the
following test, although not mandatory, is recommended to be performed annually using a calibrated
medical device safety analyzer. The procedure is as follows:
1. Disconnect the cart from mains power (unplug cart from the wall).
2. Remove power from internal power system outlets by pressing the power button on the Power System
Interface.
3. Set up the Insulation Resistance Test per the medical device safety analyzer instructions.
4. Connect the Cart power cord to the medical device safety analyzer.
5. Perform test (500 V).
6. Ensure Insulation Resistance is greater than or equal to 1 Mohms.
7. Remove medical device safety analyzer connections to Cart and return Cart to service.
3OHDVH5HDGDQG)ROORZ3URGXFW0DQXDO,QVWUXFWLRQV
THIS IS A POLARIZED CORD - HOSPITAL GRADE ONLY
127(*URXQGLQJ5HOLDELOLW\&DQ2QO\%H$FKLHYHG:KHQ
WKH(TXLSPHQW,V&RQQHFWHG7R$Q(TXLYDOHQW5HFHSWDFOH0DUNHG
³+263,7$/21/<´2U³+263,7$/*5$'(´
820-672
,QVSHFW&RUG7KRURXJKO\%HIRUH(DFK8VH'212786(,)'$0$*('
'R1RW3OXJ0RUH7KDQ63(&,),('180%(52):$776,QWR3RZHU6\VWHP
'R1RW5XQ&RUG7KURXJK'RRUZD\V+ROHVLQ&HLOLQJV:DOOVRU)ORRUV
)8//<,16(573OXJ,QWR2XWOHW
'R1RW5HPRYH%HQGRU0RGLI\$Q\0HWDO3URQJVRU3LQVRI&RUG
'R1RW8VH([FHVVLYH)RUFHWR0DNH&RQQHFWLRQV
.HHS$ZD\)URP:DWHU'212786(:+(1:(7
.HHS&KLOGUHQ$ZD\)URP&RUG
'R1RW3OXJ&RUGLQWR$Q([WHQVLRQ&RUG
$92,'29(5+($7,1*8QFRLO&RUGDQG'R1RW&RYHU,W:LWK$Q\0DWHULDO
'R1RW'ULYH'UDJRU3ODFH2EMHFWV2YHU&RUG'R1RW:DON2Q&RUG
*5$633/8*WR5(029()5202XWOHW'R1RW8QSOXJ%\3XOOLQJ2Q&RUG
DANGER!
ELECTRICAL CORDS CAN BE HAZARDOUS
Misuse Can Result in FIRE or DEATH by ELECTRICAL SHOCK.
LES CORDONS ELECTRIQUES PEUVENT ETRE DANGEREUX
Une mauvaise utilisation peut provoquer un INCENDIE ou la MORT par CHOC
ELECTRIQUE.
820-672