Technical data

Flow Meter FC01-Ex Flow Meter FC01-Ex
12 Ex atmosphere - Definitions and mounting instructions 12 Ex atmosphere - Definitions and mounting instructions
2.6.2 Installation of electronic control unit FC01-Ex
2.6.2.1 Mechanical installation
The enclosure is installed by 4 retaining screws.
The blue strips over the screws should be removed to allow the front cover to be eased
out of the enclosure.
The surface mounted enclosure meets protection degree IP54.
2.6.2 Installation of electronic control unit FC01-Ex
2.6.2.1 Mechanical installation
The enclosure is installed by 4 retaining screws.
The blue strips over the screws should be removed to allow the front cover to be eased
out of the enclosure.
The surface mounted enclosure meets protection degree IP54.
90
PG9 / M16
PG11 / M20
120
240
G13.5
(blue)
133
226
ø5
PG11 / M20
1
2
3
fig. 7
90
PG9 / M16
PG11 / M20
120
240
G13.5
(blue)
133
226
ø5
PG11 / M20
1
2
3
fig. 7
2.6.2.2 Electrical connection
Take the FC01-Ex equipotential bonding cables (1,5 mm
2
) from the monitoring head
through the cable gland 1 (fig. 7) and to the centre grounding system (fig. 6) and connect
to terminal USLKG5.
Take the cable of the monitoring head through the cable gland and connect to the barriers
according to the connection scheme FC01-Ex (fig. 6).
Take supply feed through cable gland 2 and further required connection cables through
cable gland 3 (fig. 7) and connect to terminals XV (see connection scheme fig. 6).
2.6.2.2 Electrical connection
Take the FC01-Ex equipotential bonding cables (1,5 mm
2
), from the monitoring head
through the cable gland 1 (fig. 7) and to the centre grounding system (fig. 6) and connect
to terminal USLKG5.
Take the cable of the monitoring head through the cable gland and connect to the barriers
according to the connection scheme FC01-Ex (fig. 6).
Take supply feed through cable gland 2 and further required connection cables through
cable gland 3 (fig. 7) and connect to terminals XV (see connection scheme fig. 6).