Technical data

Flow meter FC01-Ex Flow meter FC01-Ex
On-line phases 53 On-line phases 53
6 On-line phases
6.1 Switch-on performance
Upon power application, POWER ON TEST, will be shown on the display for approx. 1 sec, with
the software version number being indicated in the second line.
During this period, the integral controller will conduct test routines (see para. 7.1, Test and diagnosis).
If during the test no error was found, the display will indicate HEATING UP.
The FC01-Ex will then be in the heating up period required for the measuring procedure.
6.2 Measuring cycle
Upon completion of the heating up period and availability of the first measured value, the
display will change to measuring cycle, and the user interfaces such as analogue outputs or limit
switches will be up-dated.
The following operating data may be retrieved in the main menu during the measuring cycle:
6.2.1 Operating data
6.2.1.1 Measured value(s)
Flow rate and medium temperature are indicated by the units selected in the upper line of the
LC display.
The lower line of the display will optionally show the switching condition of the limit switches and
an analogue bar with a 10-segment resolution, or the flow volume/time unit pertinent to the indi-
cated flow rate or the totalized flow volume (totalizer function).
The analogue bar has different meanings, depending on its configuration (see para. 5.1.1.9 -
menu option BARGRAPH).
The limit switches are identified according to their physical assignment, i.e. by F for flow rate
and T for medium temperature, at the first or last place of the second line on the display.
If F and T are shown reversed, the limit switch is in the switch-on condition.
Limit switches lying within the analogue bar range are also represented at the appropriate place
of the analogue bar (see para. 5.1.1.9).
6 On-line phases
6.1 Switch-on performance
Upon power application, POWER ON TEST, will be shown on the display for approx. 1 sec, with
the software version number being indicated in the second line.
During this period, the integral controller will conduct test routines (see para. 7.1, Test and diagnosis).
If during the test no error was found, the display will indicate HEATING UP.
The FC01-Ex will then be in the heating up period required for the measuring procedure.
6.2 Measuring cycle
Upon completion of the heating up period and availability of the first measured value, the
display will change to measuring cycle, and the user interfaces such as analogue outputs or limit
switches will be up-dated.
The following operating data may be retrieved in the main menu during the measuring cycle:
6.2.1 Operating data
6.2.1.1 Measured value(s)
Flow rate and medium temperature are indicated by the units selected in the upper line of the
LC display.
The lower line of the display will optionally show the switching condition of the limit switches and
an analogue bar with a 10-segment resolution, or the flow volume/time unit pertinent to the indi-
cated flow rate or the totalized flow volume (totalizer function).
The analogue bar has different meanings, depending on its configuration (see para. 5.1.1.9 -
menu option BARGRAPH).
The limit switches are identified according to their physical assignment, i.e. by F for flow rate
and T for medium temperature, at the first or last place of the second line on the display.
If F and T are shown reversed, the limit switch is in the switch-on condition.
Limit switches lying within the analogue bar range are also represented at the appropriate place
of the analogue bar (see para. 5.1.1.9).