Technical data

Flow meter FC01-Ex Flow meter FC01-Ex
Technical data 65 Technical data 65
8.6 Metrological data
8.6.1 Mass flow measurement:
Medium: air water
Measuring range: 0 … 20 m/s 0 … 3 m/s
Accuracy: ± 10% MBE *
1)
±10% MBE *
Repeatability: ± 1% MW ** ±1% MW **
(5% MBE – 100% MBE)
8.6.2 Temperature measurement:
Measuring range: -40 … +90 °C 0 … +90 °C
Accuracy: ± 1% MB *** ± 1.5% MB ***
8.6.3 FC01-Ex Electronic control unit
Temperature drift of the electronic
control unit: ± 0.1%/K/MBE * ± 0.35%/K/MBE *
Warm-up to full accuracy: 15 min 15 min
* MBE - of final value
** MW - measured value
*** MB - measuring range
1)
Please enquire for higher accuracy