Technical data

Flow Meter FC01-Ex Flow Meter FC01-Ex
68 Examples 68 Examples
After trim point 10 has been set at 1.80 m/s and the velocity has actually been controlled accordingly,
automatic calibration is started by simultaneously pressing UP and DOWN.
The following Y value was determined for flow velocity 1.80 m/s after completion of the heating
up period and the calibration.
After trim point 10 has been set at 1.80 m/s and the velocity has actually been controlled accordingly,
automatic calibration is started by simultaneously pressing UP and DOWN.
The following Y value was determined for flow velocity 1.80 m/s after completion of the heating
up period and the calibration.
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
Trim point V [m/s]
10
9
8
7
6
5
4
3
2
1
Trim point V [m/s]
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
10
9
8
7
6
5
4
3
2
1
Y
35400
This value represents the heating power required to measure the velocity of 1.80 m/s.
After this value has been transferred, trim points 9 ... 1 should be processed the same way.
The Y values shown in the following table were measured for trim points 9 ... 1.
The medium temperature at which the calibration was made (here: TRIM IS READY!
TEMP = 82.8 °C) is indicated on the display after the last trim point.
The CUSTOMER TRIM procedure is now completed !
This value represents the heating power required to measure the velocity of 1.80 m/s.
After this value has been transferred, trim points 9 ... 1 should be processed the same way.
The Y values shown in the following table were measured for trim points 9 ... 1.
The medium temperature at which the calibration was made (here: TRIM IS READY!
TEMP = 82.8 °C) is indicated on the display after the last trim point.
The CUSTOMER TRIM procedure is now completed !
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
Trim point V [m/s]
10
9
8
7
6
5
4
3
2
1
Trim point V [m/s]
1.80
1.60
1.40
1.20
1.00
0.80
0.60
0.40
0.20
0.00
10
9
8
7
6
5
4
3
2
1
Y
35400