Technical data

Flow Meter FC01-Ex Flow Meter FC01-Ex
72 Examples 72 Examples
CHARACTERISTIC new
NUMBER OF TRIM POINTS 12
TEMPERATURE DIFFERENCE 3.3
Take the data for the curve from the table of page 71 and set them on the FC01-Ex.
Flow velocity 2.10 m/s shall be assigned to trim point 12. Other than with the automatic calibration
the applicable Y value of 35441 shall be entered by means of the keyboard. 1.95 m/s and 35396
are set for point 11 etc.
The procedure is repeated until the entire curve has been determined.
After the data for point 1 have been entered and confirmed, the display indicates TRIM IS
READY! TEMP = 25.0 °C.
Other than with the automatic calibration, the calibration temperature at which the curve has
been determined will be flashing on the display and must be set at +82.8 °C by hand (example 1
“Implementation”).
The calibration data are completely entered!
Before quitting the submenu, press M to permanently store the calibration data.
CHARACTERISTIC new
NUMBER OF TRIM POINTS 12
TEMPERATURE DIFFERENCE 3.3
Take the data for the curve from the table of page 71 and set them on the FC01-Ex.
Flow velocity 2.10 m/s shall be assigned to trim point 12. Other than with the automatic calibration
the applicable Y value of 35441 shall be entered by means of the keyboard. 1.95 m/s and 35396
are set for point 11 etc.
The procedure is repeated until the entire curve has been determined.
After the data for point 1 have been entered and confirmed, the display indicates TRIM IS
READY! TEMP = 25.0 °C.
Other than with the automatic calibration, the calibration temperature at which the curve has
been determined will be flashing on the display and must be set at +82.8 °C by hand (example 1
“Implementation”).
The calibration data are completely entered!
Before quitting the submenu, press M to permanently store the calibration data.