Model 2090 Multi-Device Controller
11. Reconnect all cables and power up the Model 2090.
After a moment, all of the indicators should flash and
then the firmware revision will be displayed in the left
display. The unit will then display
E001 error in the
left display indicating the NVRAM values have been
reset to the factory defaults. Refer to the list made in
step one, or the parameter settings in the “Model 2090
Parameter List” section, to make any necessary changes
based on your equipment configuration.
rrect
ockets (HI vs. LO) and orientation (notch indicating pin
Model 2090) and that all
M, or contact ETS for
ssistance. If the Model 2090 powers on normally but only
TROUBLESHOOTING
If your Model 2090 fails to function correctly, first check
all power and fiber connections. If the Model 2090 power
light comes on, but the display and indicators never flash
on or show the revision number, check that the SBC is
firmly seated, that the ROMs are installed in the co
s
one facing towards the front of the
IC pins are properly seated in their sockets. If the Model
2090 powers on and shows “
FLASH CODE” in the display,
this indicates that the Model 2090 is in bootstrap mode
awaiting a flash upgrade. Either the bootstrap code
detected invalid firmware upon power up, or there is a
problem with the Model 2090 front panel. Power the unit
off and back on to see if the problem persists. If so, you
can try downloading the flash upgrade file for the Model
2090 and restoring the corrupted RO
a
shows dashed lines (“
-----”) in the displays, insure that
the fiber optics to the device(s) are connected correctly