Owner's manual

www.ets-lindgren.com
Specifications
3.0 Specifications
Electrical Specifications
Frequency Range:
100 kHz65 MHz
Lines plus Ground:
2
Maximum Load Capacity:
25 A
This rating may be increased to
30 A provided that the duty cycle is
reduced to 20%.
Maximum Voltage
Line-to-Ground:
220 VAC
Maximum line-to-line voltage must
not exceed 220 volts AC RMS.
Line-to-Line:
400 VAC
Maximum line-to-ground voltage
must not exceed 440 volts AC RMS.
Network Inductance:
5 µH / 50
Power Source Frequency:
DC 400 Hz
Environmental
Installation:
Indoor use only
Altitude:
15000 ft (4572 m) max
Temperature:
0°C to 40°C (32°F to 104°F)
Relative Humidity:
80% up to 31°C (87.8°F)
decreasing linearly to 50% at
40°C (104°F)