108 | Troubleshooting and Error Handling
ERROR CODE Q: SAMPLE/S–1 PROCESS ERR–RATE TOO HIGH
A probe is connected that does not support the current sample rate of
the FM5004. The sample rate can only be set to the rate at which the
slowest probe will operate.
ERROR CODE R: RECEIVE ERR DATA NOT REC FROM PROBE
The FM5004 sent a command to a probe and no response was
received for one sample cycle.
ERROR CODE S: SEND ERROR PROBE NOT READY TO SEND
The FM5004 was ready to send a command to a probe but the
complete response was not received from the previous command.
ERROR CODE U: TEMP UPPER TEMP OVER UPPER LIMITS
A temperature measurement caused an upper temperature alarm to
occur.
ERROR CODE W: BATT WARN AT WARNING LEVEL
The battery voltage of a probe fell below the warning level. This
indicates that the probe will operate for only a short time before the
battery fails. Charge the probe battery as soon as possible.
ERROR CODE Y: ALARM UPPER 4/3 DISPLAY ALARM
A field measurement caused the 4-probe or 3-axis upper alarm to
occur.
ERROR CODE Z: ALARM LOWER 4/3 DISPLAY ALARM
A field measurement caused the 4-probe or 3-axis lower alarm to
occur.