Communication Error Codes
Appendix B: Communication Error Codes
When the HI-4416 detects an error condition during communication with a probe,
it will display an error message of the form EXX, where XX is a two-digit number.
The meaning of each error number is described below. Under certain
circumstances, it is possible for the user to correct the conditions causing errors
01 and 02: if not, or if the conditions that generate errors 03 - 12 develop and
persist, the probe must be repaired by ETS-Lindgren technicians.
Error
Cause
E01No Response From Probe
Probe's ARM/OFF switch in OFF
position; faulty probe.
E02Transmission Error (e.g., Parity)
The HI-4416 is ON, no fiber optic
cables are connected, and the
readout's connectors are aimed
toward a light source; faulty probe.
E03Input Buffer Overflow
Too many characters contained
between the Start Character/Carriage
Return sequence.
E04Invalid start character for probe
data
Start Character incorrect or not sent.
E05Probe Data String Length Error
Data string does not conform to one
of the two correct string lengths.
E06Invalid String for Reading Value
Data string does not conform to
correct format (four digits plus decimal
point).
E07Invalid Range Value
Incorrect range character.
E08--Invalid Unit Value
Incorrect unit characters.
E09Invalid Axis Enable Value
Value is other than "E" or "D".
E10Invalid Battery Status Value
Value is other than "N", "D" or "F".
E11Over Range Indicator
Value is other than "N" or "O".
E12Invalid Recorder Out Value
Value is not in the range 0 255.