6. Selecting the alarm contact type: press the or key to select the alarm contact
type:
“STDY” = steady contact
“FLET” = fleeting (single pulse) contact
Press the ENT key to confirm your selection.
Note: With this parameter you select whether the alarm contact will operate as a
steady contact or a fleeting (single pulse) contact. Pulse contact closing time is 1
second.
7. Setting the line resistance: press the or key to set the line resistance of the
sensor cable connected to the transmitter. Press the ENT key to confirm your setting.
Note: This feature will be available only if “2 cell” is selected.
8. Resetting the transmitter settings to factory defaults: The display shows “NO DEF”.
Press the or key to select:
“NO DEF” = keeps old values active, when confirmed with ENT key
“YES DEF” = resets all settings to factory defaults, when confirmed with ENT key
Press the ENT key. The selected function will be executed and the transmitter returns
automatically to the Measurement mode.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).
6.11 Calibration (CAL) sub-function
The calibration procedure in Setup mode is identical to the procedure in the Calibration mode
(see section 5). The only difference is that the transmitter remains in Setup mode (instead of
Measurement mode) after calibration is completed.