5 
2.3 Connecting Peripherals 
2.3.1 Connection Terminals 
Remove Back Cover: 
Remove the screws from the four corners at the back of the conductivity 
controller. Remove the back cover. The connectors are exposed on the 
back PCBA as shown in the Figure 1 below. 
Connectors: 
•  J11 – 24V DC power connector 
•  J8 – 9V DC power connector (1.2mm diameter) 
•  J9 – Relay connections 
•  J10 - Conductivity electrode & Temperature probe connections (wiring 
has to be done in the detachable connector 
J10
J8
REL 1
REL 2
J9
9V DC Power
Screw
Figure 1: Outer Side of Back PCBA 
J11 Connections 
1. +ve 
2. Gnd 
J10 Connections 
1.  CON Low Current 
2.  CON Low Voltage 
3. CON High Voltage 
4. CON High Current 
5.  Pt 100 Compensate 
6.  Pt 100 Sense 
7.  Pt 100 GND 
8. (No connection) 
J9 Connections 
Relay 1 & 2 
1)+ve 
2)Gnd 
24V DC 