Instruction manual
15 
4  CALIBRATION MODE 
4.1 About Calibration 
Calibration should be carried out each time a new electrode is attached to the 
controller or when you suspect that the controller/electrode is out of calibration. The 
controller allows you to perform temperature calibration and conductivity calibration. 
About Temperature Calibration 
It is important to ensure that temperature calibration is carried out prior to 
conductivity calibration since temperature readings affect the accuracy of 
conductivity measurements. It is recommended that temperature calibration should 
be carried out only if the temperature value displayed on the controller is different 
from that of a calibrated thermometer. A temperature offset calibration of ± 10 °C/± 
18 °F from the default reading is allowed. Once a temperature calibration is 
performed, conductivity calibration should be carried out to ensure the accuracy of 
conductivity measurements. 
About Measuring Ranges & Conductivity Calibration 
The controller has six selectable conductivity measuring ranges. You can only 
calibrate one-point in the measuring range that you have selected for your process 
application. You have to recalibrate the controller every time you change the 
measuring range since the calibration data of the controller will be reset every time 
a new measuring range is selected. 
About Conductivity Calibration 
During conductivity calibration, the controller allows a calibration window of ± 40% 
from the measured default reading of the calibration standard. The minimum 
allowable calibration point is 10% of the full scale reading of the measuring range 
selected. 
For best results, select a standard conductivity solution close to the sample value 
you are measuring. 










