Instruction manual
19 
5.2  Temperature Coefficient Settings 
Setting the temperature coefficient to match your process/calibration liquids allows 
controller to achieve conductivity readings with higher accuracy. 
ENT
SETUP
HOLD
1
SETUP
HOLD
2
ENT
SETUP
HOLD
3
ENT
4
ENT
SETUP
HOLD
2
%
SETUP
HOLD
%
ENT
11
From measurement mode press ENT key to enter setup mode as 
described in section 5.1. The LCD shows the first screen of setup mode 
(rAng). Press ▲ or ▼ key to select temperature coefficient settings 
screen (tC). 
Press ENT key to access temperature coefficient settings (tC). 
2
Select type of temperature coefficient: The upper display shows the 
last configured temperature coefficient type (Lin or PUr). Press ▲ or 
▼ key to select the desired type based on type of your process liquid. 
 PUr :  Pure water temperature coefficient: Select this if your 
process liquid is pure water or ultra pure water. 
 Lin :  Linear temperature coefficient: Select this for all other 
process liquids 
Press ENT key to confirm your selection. 
If PUr was selected: The controller reverts back to tC screen 
If Lin was selected: 
3
Select temperature coefficient for process liquid: The lower display 
shows ‘P.tC’. The upper display shows last configured temperature 
coefficient for process liquid (if any). The default value is 2.1%. Press 
▲ or ▼ key to select the desired value. Allowable range: 0 to 10%. 
Press ENT key to confirm your value. 
4
Select temperature coefficient for calibration liquid: The lower 
display shows ‘C.tC’. The upper display shows last configured 
temperature coefficient for calibration liquid (if any). The default value is 
2.1%. Press ▲ or ▼ key to select the desired value. Allowable range: 
0 to 10%. Press ENT key to confirm your value. The controller reverts 
to tC screen. 
Press  ▲ or ▼ key to access other setup screens or press ▲ and ▼ key 
simultaneously (escape) to return to measurement mode. 