5.5  Controller Settings 
The controller settings allow you to set the relay type and relay status of the 
controller. 
ENT
SETUP
HOLD
2
ENT
SETUP
HOLD
3
SETUP
HOLD
1
SETUP
HOLD
2
ENT
SETUP
HOLD
3
ENT
ENT
11
From measurement mode press ENT key to enter setup mode as 
described in section 5.1. The LCD shows the first screen of setup 
mode (rAng). Press ▲ or ▼ key to select controller settings screen 
(Cntr). 
Press ENT key to access controller settings (Cntr). 
2
Select the controller type: The lower display shows ‘tyP’. The upper 
display shows the last configured controller type. (‘OFF’ or ‘L.Ct’). 
Press ▲ or ▼ key to select the desired controller type. 
 OFF : 
Controller Off 
Selecting ‘OFF’, the conductivity controller stops 
functioning as a controller by preventing relay functions. 
Once the controller type is set to ‘OFF’, the setup mode 
does not show you the Relay 1/Relay 2 settings. 
 L.Ct:  Limit value control (Controller On) 
Selecting ‘L.Ct’, the conductivity controller is set to function 
as a controller. When the controller type is set to ‘L.Ct’, the 
Relay 1/Relay 2 settings are available in the setup mode 
for you to configure relay operating parameters. 
Press ENT key to confirm your selection. 