Instruction manual
26 
5.6 Measuring Range Settings 
The conductivity controller has 6 different measuring ranges. You need to select a 
range which is the most suitable for conductivity reading of your 
applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm). 
The measuring range you select here will affect the way the controller behaves in 
the measurement mode and calibration mode. In measurement mode, if the 
conductivity reading of the controller is above the selected measuring range, the 
LCD shows “Or” (over range) error. In calibration mode, the controller adjusts 
calibration tolerance based on full scale of the selected measuring range. The 
calibration is carried out only for the selected measuring range. 
WARNING: If you change the measuring range of the controller after calibration is 
done, the controller will reset back to its factory default calibration. 
Press  ▲ or ▼ key to access other setup screens or press ▲ and ▼ key 
simultaneously (escape) to return to measurement mode. 
11
From measurement mode press ENT 
key to enter setup mode as described in 
section 5.1. The LCD shows the first 
screen of setup mode (rAng). Press 
ENT key to access measurement range 
setting (rAng). 
2
The upper display shows the currently 
selected measuring range. The factory 
default is 2000 µS. The lower display 
shows the currently configured cell 
constant (k). (Refer section 4.2 for 
setting cell constant) 
Press ▲ or ▼ key to select the desired 
measuring range. Press ENT key to 
confirm your selection. 
  The controller reverts to rAng screen. 
Measuring range  Resolution 
0.000 to 2.000 µS/cm  0.001 µS/cm 
0.00 to 20.00 µS/cm  0.01 µS/cm 
0.0 to 200.0 µS/cm  0.1 µS/cm 
0 to 2000 µS/cm  1 µS/cm 
0.00 to 20.00 mS/cm  0.01 mS/cm 
0.0 to 200.0 mS/cm  0.1 mS/cm 
SETUP
HOLD
ENT
1
µS
HOLD
SETUP
2
µS
HOLD
SETUP
2
µS
HOLD
SETUP
2
µS
HOLD
SETUP
2
mS
HOLD
SETUP
2
mS
HOLD
SETUP
2
ENT
ENT
ENT
ENT
ENT
ENT










