6  TECHNICAL SPECIFICATIONS 
General Specification 
Conductivity 
Measuring Range: Resolution:
0.000 - 2.000 µS/cm  0.001 µS/cm 
0.00 - 20.00 µS/cm  0.01 µS/cm 
0.0 - 200.0 µS/cm  0.1 µS/cm 
0- 2000 µS/cm  1 µS/cm 
0.00 - 20.00 mS/cm  0.01 mS/cm 
0.0 - 200.0 mS/cm  0.1 mS/cm 
Relative accuracy  ± 1% of full scale reading 
System accuracy  The effective measuring range and the 
total system accuracy depend on the 
sensor used with the controller. Please 
refer respective sensor documentation. 
Temperature 
Measuring range  -10.0 to +125.0 
o
C 
+14.0 to +257.0 
o
F 
Resolution 0.1 
o
C / 0.1 
o
F 
Relative accuracy  ±0.5 
o
C / ±1.0 
o
F 
Sensor  2 wire / 3 wire - Pt100 or Pt1000 (Jumper 
Selectable) 
Compensation  Automatic or Manual (reference at 25
o
C) 
Calibration 
Conductivity 
Number of calibration points  1 point 
Calibration window  ±40% of default reading 
Calibration standard solution 
window 
10% to 100% of the selected measuring 
range 
Temperature 
Offset Adjustment  ± 10 ˚C/ ± 18 ˚F 
Controller 
Output 2 SPST Relays 
Relay 1/Relay 2 function (for 
Set Point 1 (SP1) & Set 
Point(SP2)) 
LO (Relay is active when measured value 
< Set point) 
HI (Relay is active when measured value > 
Set point) 
Hysteresis window  0% to 10 % of selected measuring range 
On-delay & Off-delay time  0 to 2000 seconds 
Display 
LCD  7 segments display with symbols for status 
information 
Back light  On/Off selectable 