Instruction manual
Instruction manual  αlpha DO 2000 
10 
3.2  Connection Diagram 
 Caution: Ensure electrical mains are disconnected before proceeding. 
1
L
2
N
Current
15
Current
13
OP2
14
16
OP1
17
3
4
5
6
22
HOLD
18 19 2120 23 24 25
7
8 9
10
11
PT1000
29
PT100/
26 2827 30 31 32 33
12
Fuse: 250V 315 mA
Time Delay
11
1
L
2
N
13 14
OP2
Curre nt
15 16
OP1
Curre nt
63
19
HOLD
1817 20 21
5
4
22 23 24 25
28
PT100/
7
2726
8
3029
9
PT1000
31
10
32
33
12
250 V 315 mA
Time Delay
Connections for wall mounting version 
1.  AC mains live wire  18.  12V output Power supply 
2.  AC mains neutral wire  19.  Hold Function 
3.  AC mains protective earth wire  20.  Hold Function 
4.  Relay A (SP 1)  21.  Earth ground 
5.  Relay A (SP 1)  22.  no connection 
6.  Relay B (SP 2)  23.  no connection 
7.  Relay B (SP 2)  24.  no connection 
8.  Wash relay  25.  no connection 
9.  Wash relay  26.  no connection 
10.  Alarm relay (NC)  27.  no connection 
11.  Alarm relay common  28. Temperature ground 
12.  Alarm relay (NO)  29. Temperature Sense 
13.  4-20 mA temperature output, +ve terminal 30.  Temperature Input 
14.  4-20 mA temperature output, -ve terminal    (Short to terminal 29 if using 2- wire RTD) 
15.  4-20 mA DO output, +ve terminal 31. no connection 
16.  4-20 mA DO output, -ve terminal  32.  DO input, Cathode, -ve terminal 
17.  12V output ground  33.  DO input, Anode, +ve terminal 
IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF 
under Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ 
contacts of the relay (10 & 11). 