−  If the proportional controller type is set to pulse frequency control (PFC): 
9.  Selecting the relay status under Non-Alarm condition: press the ▲ or ▼ key to choose 
the desired relay status (de-energized = “DEEN” or energized = “EN”). Press the ENT key to 
confirm your selection. 
10. Setting the proportional range: press the ▲ or ▼ key to set the proportional range (setting 
range: 10 to 500%). Press the ENT key to confirm your setting. 
11. Setting the pulse frequency: press the ▲ or ▼ key to set the pulse frequency (setting 
range: 60 to 120 pulses per minute). Press the ENT key to confirm your setting. 
12. Setting the integral action time: press the ▲ or ▼ key to set the integral action time 
(setting range: 0.0 to 999.9 minutes). Press the ENT key to confirm your setting. 
Continue with Setup mode procedures, or return to Measurement mode by pressing the ▲ 
and ▼ keys simultaneously (escape). 