Instruction manual
Instruction manual  αlpha DO 2000 
32 
8  TECHNICAL SPECIFICATIONS 
8.1  Measurement and Compensation
(a) Dissolved Oxygen 
Dissolved Oxygen Measuring Range  0.00 to 25.00 mg/l or 0.00 to 25.00ppm or 
00.0 to 300.0 % Saturation 
Resolution  0.01 mg/l or 0.01 ppm or 
0.1 % 
Relative accuracy  1.5 % of full-scale reading for all ranges 
(b) Temperature 
Temperature measuring range  -10.0 to +125.0 
o
C or 
+14.0 to +257.0 
o
F 
Resolution 0.1 
o
C or 
0.1 
o
F 
Relative accuracy  0.5 
o
C or 
1.0 
o
F 
Sensor  Pt100 or Pt1000 (Jumper Selectable) 
– 2 wire or 3 wire 
(c) Compensation 
Temperature Compensation Automatic or Manual 
Pressure Compensation  0.740 to 3.000 BAR or 555 to 2250 mmHg 
or 10.73 to 43.51 PSI 
(Manual setting and automatic correction) 
Salinity Compensation  0.0 to 50.0 ppt 
(Manual setting and automatic correction) 
8.2  Set-point and control function – Relay 1 & Relay 2 
Function Switchable 
- Selection in Control (CNTR) function of 
Setup 
Off 
Limit Control 
P/PI Control (Pulse length / Pulse frequency) 
(a) Off 
The control function of the controller is off  SP1 & SP2 function will not appear in setup when Off is 
selected in CNTR function to off mode 
(b) Limit Control & P/PI Control 
Relay contact selection for active mode  Energise (EN) or 
De-energise (DEEN) 
Set Point 1 (SP1)  0.0 to 300.0 % (0.00 to 25.00 mg/l) (0.00 to 25.00 ppm) 
Set Point 2 (SP2)  0.0 to 300.0 % (0.00 to 25.00 mg/l) (0.00 to 25.00 ppm) 
Active mode selection (for SP1 & SP2)  LO (Relay is active when measured value < set point) 
HI (Relay is active when measured value > set point) 
Hysteresis (for SP1 & SP2)  1.0 to 10.0 % 
On (pickup) Delay (for SP1 & SP2)  0 to 2000 seconds 
Off (Dropout) Delay (for SP1 & SP2)  0 to 2000 seconds 