P2.3: Set Hysteresis value
Hysteresis prevents rapid contact switching if your value is fluctuating near the set
point. Once activated, the relay will not return to resting position until the measured
value has passed through the set point plus hysteresis value.
Example: With a high set point of 10.00 and a hysteresis value of 0.5, the relay will
activate at pH values above 10.00, but will not de-activate till the measured pH
value decreases below pH 9.50.
The hysteresis window can be set to any value within the range shown below:
pH mode: 0.00 to 2.00 pH
ORP mode: 0 to 300 mV
1. Key in the password “022” as per procedure in Section 5.
2. Press key once. The screen will scroll P2.0 and SP2. Press ENTER three
times.
3. The screen will scroll P2.3, HYS2, and current hysteresis value.
4. Press the or keys to adjust the hysteresis value.
5. Press ENTER to confirm and return to general Setup mode.
Press and keys together (ESCAPE) to return to measurement mode.