User Guide

- 23 -
P3.3: Select temperature compensation
This controller features selectable automatic temperature compensation (ATC) or
manual temperature compensation (MTC). ATC automatically compensates for
temperature fluctuations; MTC let you select a specific value at which temperature
will be compensated. ATC readings require a temperature sensor.
1. Key in the password “022” as per procedure in Section 5.
2. Press key twice. The screen will scroll P3.0 and CONF. Press ENTER
three times.
3. The screen will scroll P3.3 and MTC (or ATC).
4. Press or keys to toggle between MTC and ATC.
5. Press ENTER to confirm and return to general Setup mode.
Press and keys together (ESCAPE) to return to measurement mode.
NOTE: ORP is not affected by temperature and so ATC does not operate in ORP
mode. When set for ATC, the pH 190 will display temperature if a 100 Pt 100
sensor is connected.