Instruction manual

Instruction Manual αlpha-pH2000D
- 27 -
6.7 Current Output 2 sub-function
In this sub-function you set the current output range of the transmitter for temperature values.
ENT
ENT
ENT
ENT
ENT
ENT
ENT
o
c
o
c
o
c
o
c
2
2
3
3
4
4
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
1
cur.2
SETUP
HOLD
1. Select the “CUR.2” sub-function, then press the ENT key.
2. Selecting the output type: press the or key to select the desired output type: 0-20 mA
or 4-20 mA. Press the ENT key to confirm your selection.
3. Setting the temperature value at which the transmitter output will be 4 mA (or 0 mA):
press the or key to set the temperature value (setting range: -10.0 to 125.0 °C or 14 to
257 °F for pt100/1000, -10.0 to 110.0 °C or 14 to 230 °F for NTC300 thermistor) to be
equivalent to 4 mA (or 0 mA). Press the ENT key to confirm your setting.
4. Setting the temperature value at which the transmitter output will be 20 mA: press the
or key to set the temperature value (setting range: 0.0 to 110.0 °C or 32 to 257. °F for
pt100/1000, 0.0 to 110.0 °C or 32 to 230 °F for NTC300 thermistor) to be equivalent to 20 mA.
Press the ENT key to confirm your setting.
Continue with Setup mode procedures, or return to Measurement mode by pressing the
and keys simultaneously (escape).