6 TDS CALIBRATION
IMPORTANT: When Calibration mode is entered, controller automatically goes into
a “HOLD” mode where the 4-20 mA output freezes and relays are de-activated (if it
was in an activated condition). Upon return to measurement mode, both 4-20mA
output and relay activities resume, depending on settings.
The αlpha TDS 190 controller includes 8 TDS measurement ranges. One-point
slope calibration is possible, in each range.
Choose fresh standard solutions, whose value is reasonably close to the
measurement value.
Before calibration, make sure electrode is clean. Use iso-propyl alcohol followed by
thorough rinsing in distilled water to clean electrode.
Important: To achieve a successful TDS calibration, two conditions must be
satisfied:
1. Difference between measured value of standard solution, and actual
value of standard solution, must be within ± 40 %.
(e.g. if value of standard solution is 1500 ppm, then allowable
measured values by controller must be within 2100ppm {+ 40%} and
900 ppm {- 40%}); and
2. The lowest possible value must be at least 10% of full scale
(e.g. if range of 0 to 100.0 ppm selected, min. measured value at
calibration, must be at least 10.0 ppm).
If any or both conditions are not satisfied, controller will display ERR 1 (blinking).
1. Press CAL / ENTER key to directly access the Calibration mode when the
unit is in conductivity measurement mode. Controller displays the value of
the standard solution.
NOTE: Instead of pressing the CAL/ENTER key, you can also enter the Calibration
mode using the password code procedure as explained in Section 5.
2. Use electrode to agitate standard solution to obtain a homogeneous solution
and to dislodge any bubbles. Allow electrode time to stabilize with solution
temperature.
3. Use / keys to adjust displayed readings to the value of the standard
solution. Press ENTER.