The temperature coefficient is a value that reflects the degree to
which the sample’s conductivity is effected by temperature changes.
The meter will use the set temperature coefficient along with the
measured temperature provided by the ATC probe to calculate what
the conductivity would be at the selected reference temperature. If no
ATC probe is being used, the meter will use the set default
temperature in the calculation.
To Set Temperature Coefficient
1. Access the Set temperature Coefficient screen from the
conductivity Setup screen. The current Temperature Coefficient
is displayed on the screen.
2. Touch clear to delete the current temperature coefficient.
3. Touch C, F or K to select the desired temperature units for the
temperature coefficient.
4. Use the keypad to enter the desired temperature coefficient.
5. Touch enter to accept the temperature coefficient and return to
the conductivity Setup screen.
OR
Touch exit to return to the Conductivity Setup screen, without
making any changes.