Replatinization is the process of replacing the platinum on the
surfaces of the 2-cell conductivity probes that may flake or wear off
over time. The platinum on the surface of the probe is used to
increase the surface area of the measuring surface resulting in
decreased polarization error. Replatinization is a relatively quick
procedure to perform, taking no more than 5 minutes in most cases.
This meter is capable of performing replatinization from 1 to 30
minutes. Although replatinization is not a long process, the
replatinization solution is costly. If you are not certain that
replatinization will improve the performance of your conductivity
probe, please call the technical support hotline.
To Perform Replatinization
1. Access the Perform Replatinization screen from the conductivity
Setup screen. The current replatinization time is displayed on the
screen.
2. Touch clear to delete the current replatinization time.
3. Use the keypad to enter the desired replatinization time.
4. Touch enter to accept the replatinization time.
5. Touch START on the display screen to begin the process of
replatinization. At this time a countdown timer will appear on the
display screen to let you know how much time is remaining in the
replatinization procedure. At anytime after you have initiated the
replatinization, you can touch STOP to abort the procedure and
return to the conductivity Setup screen.
OR
Touch exit on the display screen to cancel the replatinization
process and return to the Conductivity Setup screen.
OR
Touch EXIT to return to the Conductivity Setup screen, without
performing replatinization.
4-cell conductivity probes do not require replatinization.