Once an exact cell constant has been established, the meter is ready
for sample measurements.
1. Immerse the conductivity cell and ATC probe into the sample
solution. The immersion depth should be enough to cover the
vent holes on the electrode.
2. Stir the solution with the electrode briefly and then let the
electrode sit still in the solution.
3. Observe and record the measurement value when it appears to
be stable (STABLE will appear). The meter will automatically
change ranges from uS/ cm to mS/ cm as required.
4. Touch print to send the data to a printer or computer. If you have
assigned a sample ID# to this sample, the data will be saved in
the meter’s data storage center when you touch print.
5. Repeat steps 1 through 4 for additional samples.