User Guide

Instruction Manual CON 510
14
7 SETUP FUNCTIONS
The setup mode allows you to customize the meter’s setting to your individual preferences. The CON510 meter features
different main program menus and sub-menus which organize individual parameters like a matrix-format.
Details of each main program menu:
1. P1.0: Viewing previous calibration data - all calibrated solutions for each range
2. P2.0: Viewing electrode diagnosis - effective cell constants for each range
3. P3.0: Meter configuration – Ready, Auto-Hold, Temperature unit of measure, ATC or MTC, TDS Factor
4. P4.0: Setting of Temperature Coefficient, Normalization Temperature
5. P5.0: Setting of Automatic or Manual and Single or Multi-point calibration
6. P6.0: Setting of cell constant
7. P7.0: Resetting meter to factory defaults
7.1 SETUP Mode Overview
Press SETUP to enter into the setup mode with [SETUP] indicator on the top left of LCD display. Use the MI/ and
MR/ keys to scroll up and down respectively through main menus. At each main menu, press ENTER to enter into a
particular sub-menu to make specific changes from the meter’s default settings. See Addendum 4 for a table of meter
factory default settings.
SETUP
.
P1.0: Viewing Previous Calibration Data
P1.1 First range calibration solution
P1.2 Second range calibration solution
P1.3 Third range calibration solution
P1.4 Fourth range calibration solution
P1.5 Fifth range calibration solution
SETUP
.
P2.0: Viewing Calibrated Cell Constants
P2.1 Effective cell constant for first range
P2.2 Effective cell constant for second range
P2.3 Effective cell constant for third range
P2.4 Effective cell constant for fourth range
P2.5 Effective cell constant for fifth range
SETUP
.
P3.0: Meter Configuration
P3.1 READY indicator On/Off and Auto-Hold On
P3.2 Select °C or °F
P3.3 Select Automatic or Manual Temperature Compensation
P3.4 Setting TDS conversion factor (available in TDS mode)
SETUP
.
P4.0: Temperature
P4.1 Adjusting temperature coefficient
P4.2 Adjusting normalization temperature