7.6 P5.0 Mode of calibration
The CON 510 bench meter allows you to perform automatic (only for Conductivity
mode) or manual calibration (applicable to both Conductivity/TDS mode). Similarly you
may choose to perform either single-point or multi-point calibration (up to 4 preset
calibration standards in automatic calibration mode). Refer to the Section 4 for details
on calibration.
1. From measurement mode, press SETUP to enter into setup [SETUP] mode.
2. Use MI/ or MR/ key to scroll through main-menu “ACAL P5.0”.
3. Press ENTER key to enter into this main-menu for selection of Automatic
Calibration either “nO” (Manual) or “YES” (Automatic) using MI/ or MR/ key.
4. Once the option is selected, press ENTER key to confirm. The next display shows
the option of Single-point Calibration (SPC) or Multi-point Calibration.
5. Use MI/ or MR/ key to select “YES” if you wish to perform Single-point
Calibration and “nO” to perform Multi-point Calibration.
6. Press ENTER key to confirm after appropriate selection is made and to return to
the SETUP main-menu P5.0. Press CAL/MEAS key to return to measurement
[MEAS] mode.
.