User Guide

Instruction Manual CON 510
29
16 ADDENDUM 4: METER FACTORY DEFAULT SETTINGS
Type Parameter Default Remarks
P1.1 Viewing calibration solution –– No calibration data for first range
P1.2 –– No calibration data for second range
P1.3 –– No calibration data for third range
P1.4 –– No calibration data for fourth range
P1.5 –– No calibration data for fifth range
P2.1 Viewing probe data 1.000 No offset for effective cell constant (first range)
P2.2 1.000 No offset for effective cell constant (second range)
P2.3 1.000 No offset for effective cell constant (third range)
P2.4 1.000 No offset for effective cell constant (fourth range)
P2.5 1.000 No offset for effective cell constant (fifth range)
P3.1 Ready indicator / Auto-hold READY/ON Ready indicator on; Auto-Hold off
P3.2 Select °C/°F °C Degrees Celsius
P3.3 ATC on or off ATC YES ––
P3.4 TDS factor 0.5 Adjustable from 0.4 to 1.0
P4.1 Temperature coefficient 2.1 % per ° C Adjustable from 0 to 10%
P4.2 Normalization temperature 25° C Adjustable from 15 to 30° C
P5.0 Auto or manual calibration Auto YES Auto or manual
P5.1
Single or multi-point
calibration
Single-point YES Single or Multi-point calibration
P6.0 Cell constant 1.0 Select from k=0.1, 1.0 or 10
P7.0 Factory default No Retains your current settings