7 MEMORY FUNCTION
The DO 100 can store up to 16 sets of data. Each data
stored can either be in % Saturation, mg/L or ppm together
with its respective temperature reading taken at that
moment.
Stored data can be easily recalled based on LIFO (Last-In-
First-Out) system.
7.1 Data Input
To clear all data from memory, refer to Sections 8.1.1 (page
23) or 8.2.1 (page 31) on the procedure.
1. Switch on the meter by pressing ON/OFF key.
2. Ensure the meter is in the appropriate measurement
mode, either mg/L or % Saturation. Press MODE key
for the desired measurement mode.
3. Place the probe into the sample to be measured.
4. After the reading is stabilised, indicated by the READY
annunciator, press MI/ key. The MEM annunciator
will flash with a number indicating that the data has
been stored. See figure on right.
5. Place the probe into another sample and repeat from
step 2 above.
NOTE: The above illustration will be used in the next
Section, Memory Recall.
NOTE:
The number counter below MEM annunciator will increase
to 16 according to the following symbols:
Figure 13 Memory Input