8.1.3 P 2.2: bAr
This menu allows you to do two things:
Select the appropriate unit of measurement of pressure --- either mm Hg or
kPA.
Keying of actual atmospheric pressure value based on the altitude of the
location.
This is an important feature as DO is dependent on pressure (refer Section 12.1 on page
48).The DO 100 will use this input to compensate for the measured value automatically
without having to refer to tables.
1. Once you come to the SETUP menu shown in Figure 19 on the following page, press
ENTER.
2. Press MI/ or MR/ key to select “Hg” or “PA” as unit of pressure measurement.
3. Press ENTER key to go into the next level.
4. Press MI/ or MR/ key to adjust value accordingly to the location’s actual atmospheric
pressure.
NOTE: The range for atmospheric correction ranges from 555 to 808 mm Hg or
74 to 108 kPA.
5. Press ENTER key to confirm value.
6. The meter will accept the value and move on to the next menu.
NOTE: Press HOLD key to skip any part of the operation and move on to the
next menu, or press CAL/MEAS key to exit setup menu to go back to
measurement mode.