User Guide

Instruction Manual DO 100
31
8.2 SET Function in mg/L or ppm Measurement Mode
This part of the SETUP function allows display of values and the changing of parameters
unique to the mg/L measurement mode. Figure 25 on the following page shows the flow
diagram of the SETUP menu.
To access SETUP menu:
1. Switch on the meter.
2. Press MODE key to select mg/L.
3. Press SET key to enter the first level in its setup menu.
4. To move from one menu to the next without changing
any parameter, simply press HOLD key as shown in
Figure 25. Press ENTER or MI/ or MR/ key to
enter the menu to change parameter or simply to view
the value.
5. At any level, press CAL/MEAS to exit setup function
and go back to the measurement mode.
8.2.1 P 1.0: Clr
This clears all memory data input keyed during
measurement. Default setting is OFF.
1. Press MI/ or MR/ key to switch to “ON”.
2. Press ENTER key to confirm deleting all memory data.
3. The meter will pause briefly and move on to the next
menu.
See Figure 24.
NOTE: Press HOLD key to skip this operation and
move on to the next menu, or press
CAL/MEAS key to exit setup menu to go back
to measurement mode.
NOTE: The data cleared relates to the input taken
while the meter was in mg/L or ppm
measurement mode. To clear data input in %
Saturation measurement mode, switch the
measurement mode to % Saturation and refer to Section 8.1.1 on page
23.
Figure 24: Clear Menu